FDA Adverse Event Malfunction Summary report: N

2.2 SENO ADVANTAGE WORKSTATION

MDR report key: 1592825 · Received January 26, 2010

Report

Report Number
9611343-2010-00004
Event Type
Malfunction
Date Received
January 26, 2010
Date of Event
January 13, 2010
Report Date
January 13, 2010
Manufacturer
GE MEDICAL SYSTEMS SCS
Product Code
LLZ
PMA / PMN Number
K033400
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

AN INVESTIGATION IS ONGOING.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE 2.2 SENO ADVANTAGE SYSTEM PERFORMED A TRUE SIZE PRINTING WITH INCORRECT SCALE ANNOTATION. A GE CLINICAL SPECIALIST WAS CONDUCTING TRAINING FOR 2.2 SENO ADVANTAGE SYSTEM AT A CUSTOMER SITE. DURING THE TRAINING, THE SPECIALIST SELECTED TRUE SIZE OPTION TO PRINT THE IMAGE. THE IMAGE ON THE SCREEN WAS ANNOTATED TO A SCALE OF 1.00. WHEN THE IMAGE WAS PRINTED USING THE PRINT ORIGINAL TOOL, THE SPECIALIST NOTICED THAT THE SCALE ANNOTATED ON THE HARD COPY WAS 1.67. THERE WAS NEITHER PT INVOLVEMENT NOR INJURY REPORTED. THE CONCERN IS THAT HEALTHCARE PROFESSIONALS COULD MAKE INCORRECT MEASUREMENTS BASED ON THE SCALE WHEN PLANNING A SURGERY OR BIOPSY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 2.2 SENO ADVANTAGE WORKSTATION LLZ GE MEDICAL SYSTEMS SCS

Patients

Seq Age Sex Outcome Treatment
1