FDA Adverse Event Malfunction Summary report: N

OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM

MDR report key: 15921546 · Received December 5, 2022

Report

Report Number
3004464228-2022-22766
Event Type
Malfunction
Date Received
December 5, 2022
Date of Event
November 10, 2022
Report Date
November 10, 2022
Manufacturer
INSULET CORPORATION
Product Code
LZG
UDI-DI
10385082000009
PMA / PMN Number
K192659
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE PDM WAS RETURNED AND A VISUAL INSPECTION WAS PERFORMED; NO ABNORMALITIES WERE OBSERVED. A TECHNICAL ASSESSMENT OF THE DEVICE DESIGN IDENTIFIED THE ROOT CAUSE OF THE BATTERY EVENT TO BE THE DASH PDM CHARGING VOLTAGE EXCEEDING THE BATTERY SPECIFICATION, DEFINED AS OVERCHARGING. FIELD SAFETY CORRECTIVE ACTION HAS BEEN INITIATED BY INSULET CORPORATION. LOT RELEASE RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA.CORRECTION TO D(4): SEQUENCE NUMBER CHANGED FROM UNAVAILABLE TO 010300-11655. CORRECTION TO H(3) :DEVICE RETURNED TO MANUFACTURER CHANGED FROM NO TO YES. DEVICE EVALUATED BY MANUFACTURER UPDATED CHANGED FROM NO TO YES. CORRECTION TO H(6) INVESTIGATION FINDINGS UPDATED TO C0201. INVESTIGATION CONCLUSIONS UPDATED TO D0108.

Additional Manufacturer Narrative · 0

A TECHNICAL ASSESSMENT OF THE DEVICE DESIGN IDENTIFIED THE ROOT CAUSE OF THE BATTERY EVENT TO BE THE DASH PDM CHARGING VOLTAGE EXCEEDING THE BATTERY SPECIFICATION, DEFINED AS OVERCHARGING. FIELD SAFETY CORRECTIVE ACTION HAS BEEN INITIATED BY INSULET CORPORATION. LOT RELEASE RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA.

Description of Event or Problem · 0

IT WAS REPORTED BY THE PATIENT THAT THE OMNIPOD PERSONAL DIABETES MANAGER (PDM) IS LOSING A BATTERY CHARGE QUICKER THAN EXPECTED. THE PATIENT REPORTED THAT THEIR BATTERY WAS BULGING FROM THE BACK OF THE PDM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1947134 OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 18239 L000323 10385082000009

Patients

Seq Age Sex Outcome Treatment
1 50 YR Female