FDA Adverse Event Malfunction Summary report: N

COBAS TAQMAN 48 ANALYZER

MDR report key: 1591560 · Received February 2, 2010

Report

Report Number
2243471-2010-00003
Event Type
Malfunction
Date Received
February 2, 2010
Date of Event
January 3, 2010
Report Date
January 3, 2010
Manufacturer
ROCHE DIAGNOSTICS LTD.
Product Code
JJF
PMA / PMN Number
K012966
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS ISSUE IS CURRENTLY UNDER INVESTIGATION. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. (B)(4)

Additional Manufacturer Narrative · 1

(B)(4) COMPUTER SOFTWARE PERFORMANCE TESTS CONDUCTED, ALGORITHM ANALYSIS, SOFTWARE / FIRMWARE CAUSED EVENT, UNUSUAL EVENT. THE INVESTIGATION DETERMINED THAT THE DISCREPANT TARGET NOT DETECTED (TND) RESULT WAS CAUSED BY THE BASELINE NORMALIZATION CALCULATION OF THE ELTH ALGORITHM. IN VERY RARE INSTANCES, THE ANALYSIS CURVES OF THE AFFECTED SAMPLES ARE NOISY IN THE FIRST FEW CYCLES. THE (B)(4) CURVES DID NOT HAVE ENOUGH VALID BASELINE POINTS FOR A CORRECT CALCULATION OF THE BASELINE. THEREFORE, THE DATA POINTS OF THE PLATEAU REGION WERE INCORRECTLY USED FOR THE CALCULATION OF THE BASELINE, LEADING TO AN INCORRECT TND RESULT. ALL AFFECTED ROCHE MOLECULAR SYSTEMS ASSAYS WERE ASSESSED FOR RISKS AND POSSIBLE IMPACTS DUE TO THE ELTH BASELINE NORMALIZATION ANOMALY. ALL POTENTIAL FAILURE MODES WERE ANALYZED. THE ANOMALY HAS AN EXTREMELY LOW RATE OF OCCURRENCE. (B)(4). ALL RESIDUAL RISKS IDENTIFIED WERE DETERMINED TO BE IN THE ACCEPTABLE OR ALARP (AS LOW AS REASONABLY PRACTICAL) RANGE. THE OVERALL RESIDUAL RISK WAS DETERMINED TO BE ACCEPTABLE WITH NO MAJOR IMPACT ON ASSAY OR SYSTEM PERFORMANCE EXPECTED. (B)(4).

Description of Event or Problem · 1

A CUSTOMER IN (B)(6) REPORTED THAT THEY RECEIVED A FALSE NEGATIVE PATIENT SAMPLE RESULT FOR (B)(6) WHILE USING THE KIT C-TAQ (B)(6) 48 TESTS EXPT-IVD CE( P/N 3500756190) WITH THE COBAS TAQMAN 48 (P/N 032793320010 (B)(6), AMPLILINK V 3.2.2 (P/N 05021357001). THE RESULT THAT THE SOFTWARE REPORTED WAS "TARGET NOT DETECTED". HOWEVER, THE FLOURESCENCE AMPLIFICATION CURVE SHOWS THAT THE SAMPLE CURVE HAD POSITIVE GROWTH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COBAS TAQMAN 48 ANALYZER ANALYZER, CHEMISTRY, MICRO, FOR CLINICAL USE JJF ROCHE DIAGNOSTICS LTD.

Patients

Seq Age Sex Outcome Treatment
1