DIMENSION® CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 2517506-2010-00010
- Event Type
- Malfunction
- Date Received
- February 1, 2010
- Date of Event
- January 14, 2010
- Report Date
- January 21, 2010
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC
- Product Code
- MMI
- PMA / PMN Number
- K010313
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY DEPRESSED TROPONIN I RESULT IS UNKNOWN. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS.
SIEMENS HEALTHCARE DIAGNOSTICS FILED THE ORIGINAL MDR (B)(4) 2010. NEW INFORMATION PROVIDED BY SIEMENS HEALTHCARE DIAGNOSTICS INC. FIELD SERVICE ENGINEER (FSE) (B)(4) 2010: ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT WAS HM CHROME CONTAMINATION. THE SIEMENS HEALTHCARE DIAGNOSTICS INC. FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE ACCOUNT AND PERFORMED REPAIRS AND DE-CONTAMINATION PROCEDURES. AN MDR #1226181-2012-00023 WILL BE FILED VERSUS THE INSTRUMENT. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
A FALSELY DEPRESSED TROPONIN I RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THE SAMPLE WAS REPEATED AND A HIGHER RESULT WAS OBTAINED. IT IS UNKNOWN IF PATIENT TREATMENT WAS ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY DEPRESSED TROPONIN I RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION® CLINICAL CHEMISTRY SYSTEM | TROPONIN I FLEX® REAGENT CARTRIDGE | MMI | SIEMENS HEALTHCARE DIAGNOSTICS INC | EA0308 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |