BD INSYTE¿, PERIPHERAL IV CANNULA, BD VIALON¿ BIOMATERIAL
Report
- Report Number
- 8041187-2022-00695
- Event Type
- Malfunction
- Date Received
- December 2, 2022
- Date of Event
- November 8, 2022
- Report Date
- January 10, 2023
- Manufacturer
- BECTON DICKINSON MEDICAL (SINGAPORE)
- Product Code
- FOZ
- UDI-DI
- 30382903812234
- PMA / PMN Number
- K151698
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PO
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL YES. D10: RETURNED TO MANUFACTURER ON: 14-DEC-2022. H6: INVESTIGATION SUMMARY OUR QUALITY ENGINEER INSPECTED THE 1 PHOTO AND 35 REPRESENTATIVE SAMPLES SUBMITTED FOR EVALUATION. THE REPORTED ISSUE OF LEAKAGE WAS NOT CONFIRMED UPON INSPECTION AND TESTING OF THE SAMPLES. ANALYSIS OF THE SAMPLES SHOWED THAT THERE WERE NO ABNORMALITIES OR DAMAGES. NO ABNORMALITIES OR DAMAGES WERE OBSERVED IN THE SUPPLIED SAMPLE PHOTO AS WELL. THE RETURNED SAMPLES UNDERWENT LEAKAGE TESTING, AND THE SAMPLES SHOWED NO SIGNS OF LEAKAGE. BD COULD NOT DETERMINE A MANUFACTURING RELATED ROOT CAUSE SINCE THE DEFECT WAS NOT CONFIRMED. DEVICE HISTORY RECORD OF PACKAGED NEEDLE (PN) BATCH 2022722 CATALOGUE NUMBER 381223 AND ITS ASSEMBLED NEEDLE (AN) BATCHES 2025485 AND 1274263, PART NUMBER YF01223SGT WAS REVIEWED.
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE BD INSYTE¿, PERIPHERAL IV CANNULA, BD VIALON¿ WAS LEAKING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM PORTUGUESE TO ENGLISH: SOME CATHETERS ARE LEAKING, I.E. ALLOWING BLOOD AND CONTRAST TO PASS THROUGH. NEW INFORMATION RECEIVED: "REGARDING WHAT YOU ASK FOR REASONS I THINK ARE EVIDENT I CANNOT PHOTOGRAPH THE BLOOD COMING OUT OF THE THREAD, I CAN ONLY SAY THAT IT IS AFTER INJECTION OF CONTRAST, THE NUMBER EXACTLY, I DO NOT KNOW BECAUSE WE DO NOT COUNT, WE ARE MORE THAN ONE TECHNICIAN AND IT HAPPENED TO ALL ALWAYS WITH THE SAME (BLUE), WE OPENED THE OTHER BOX AND THEN I CAN GUARANTEE THAT IT HAPPENED TO ME 2X BECAUSE I USE THIS SIZE VERY RARELY. THIS IS THE PLACE WHERE CONTRAST ALMOST ALWAYS COMES OUT. (TALKING ABOUT THE PICTURE, THAT IS MARKED).
IT WAS REPORTED THAT THE BD INSYTE¿, PERIPHERAL IV CANNULA, BD VIALON¿ WAS LEAKING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM PORTUGUESE TO ENGLISH: SOME CATHETERS ARE LEAKING, I.E. ALLOWING BLOOD AND CONTRAST TO PASS THROUGH. NEW INFORMATION RECEIVED: "REGARDING WHAT YOU ASK FOR REASONS I THINK ARE EVIDENT I CANNOT PHOTOGRAPH THE BLOOD COMING OUT OF THE THREAD, I CAN ONLY SAY THAT IT IS AFTER INJECTION OF CONTRAST, THE NUMBER EXACTLY, I DO NOT KNOW BECAUSE WE DO NOT COUNT, WE ARE MORE THAN ONE TECHNICIAN AND IT HAPPENED TO ALL ALWAYS WITH THE SAME (BLUE), WE OPENED THE OTHER BOX AND THEN I CAN GUARANTEE THAT IT HAPPENED TO ME 2X BECAUSE I USE THIS SIZE VERY RARELY. THIS IS THE PLACE WHERE CONTRAST ALMOST ALWAYS COMES OUT. (TALKING ABOUT THE PICTURE, THAT IS MARKED).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1946019 | BD INSYTE¿, PERIPHERAL IV CANNULA, BD VIALON¿ BIOMATERIAL | INTRAVASCULAR CATHETER | FOZ | BECTON DICKINSON MEDICAL (SINGAPORE) | 381223 | 2022722 | 30382903812234 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |