FDA Adverse Event Malfunction Summary report: N

CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V

MDR report key: 15910384 · Received December 2, 2022

Report

Report Number
2249723-2022-03087
Event Type
Malfunction
Date Received
December 2, 2022
Date of Event
November 21, 2022
Report Date
March 30, 2023
Manufacturer
DATASCOPE CORP. - MAHWAH
Product Code
DSP
UDI-DI
10607567112541
PMA / PMN Number
K063525
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF OUR INVESTIGATION.

Additional Manufacturer Narrative · 0

A GETINGE FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE UNIT. THE FSE STATED THAT THE UNIT WAS NOT TURNING ON WITH ERROR # 5. IN ADDITION, THE LCD WAS NOT WORKING. THE FSE INSPECTED THE LCD AND FOUND THAT THE DISPLAY CABLE WAS NOT CONNECTED PROPERLY TO THE VIDEO RECEIVER PCB. THE FSE REINSTALLED AND CHECKED ALL CONNECTIONS ON THE MONITOR. THE FSE REPLACED THE DISPLAY TO VIDEO RECEIVER BOARD CABLE (0012-00-1429), COLOR VIDEO RECEIVER PCB (0670-00-0736), AND THE MAIN BOARD PCB (0670-00-0788). THE UNIT WAS TESTED AND FOUND TO BE WORKING PROPERLY. THE ELECTRICAL SAFETY TESTED PASSED. THE IABP WAS RETURNED TO THE CUSTOMER AND CLEARED FOR CLINICAL USE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) PANEL SHOWING DISTORTION. THERE WAS NO PATIENT HARM REPORTED.

Description of Event or Problem · 0

N/A.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING ROUTINE CHECK THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) PANEL SHOWING DISTORTION. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1960710 CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL DSP DATASCOPE CORP. - MAHWAH 0998-00-3023-55 10607567112541

Patients

Seq Age Sex Outcome Treatment
1 Unknown N/A.| UNKNOWN.