FDA Adverse Event Malfunction Summary report: N

IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 8MM

MDR report key: 15909313 · Received December 2, 2022

Report

Report Number
0002023141-2022-02997
Event Type
Malfunction
Date Received
December 2, 2022
Date of Event
September 26, 2022
Report Date
April 7, 2023
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024020085
PMA / PMN Number
K061410
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: H6: TYPE OF INVESTIGATION CODES WERE UPDATED: 10, 3331, 4110, 4109, 4111. H6: INVESTIGATION FINDINGS CODE WAS UPDATED: 213. H6: INVESTIGATION CONCLUSION CODE WAS UPDATED: 4310. ONE (1) IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 8MM (TSVWB8) WAS RETURNED FOR INVESTIGATION. VISUAL EVALUATION OF THE AS RETURNED PRODUCT IDENTIFIED SIGNS OF USE. THE REPORTED EVENT WAS UNCONFIRMED FOLLOWING FUNCTIONAL TESTING. THE COOB WAS ALSO UNCONFIRMED. BASED ON THE EVALUATION, DEVICE MALFUNCTION DID NOT OCCUR. ADDITIONALLY, THERE IS NO EXISTING ACTIONABLE TREND OR NON-CONFORMANCE / CAPA / HHE/D / IE / PRODUCT HOLD AGAINST THE REPORTED DEVICE THAT DID OR COULD CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. MONTHLY POST MARKET TRENDING REVIEW IDENTIFIED NO ACTIONABLE TRENDS OR CORRECTIVE ACTIONS FOR THE REPORTED EVENT OR DEVICE. ZIMVIE QUALITY MANAGEMENT SYSTEM (QMS) HAS CONTROLS IN PLACE TO ENSURE THE DISTRIBUTION OF CONFORMING PRODUCT. THEREFORE, BASED ON THE AVAILABLE INFORMATION, THE PRODUCT WAS WITHIN SPECIFICATIONS AND LIKELY CONFORMING WHEN IT LEFT ZIMVIE. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1243473). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1243473) FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. THE COMPLAINT IS RELATED TO THE FUNCTIONAL PERFORMANCE OF THE DEVICE. A DEFINITIVE ROOT CAUSE COULD NOT BE IDENTIFIED. HOWEVER, BASED ON THE INVESTIGATION, RISK REVIEW AND IFU, THE MOST LIKELY CAUSES DETERMINED FROM THE INVESTIGATION IS DUE TO THE TORQUE APPLIED DURING PLACEMENT/SEATING EXCEEDS RECOMMENDED VALUE, INCORRECT ASSEMBLY OF THE ABUTMENT/MOUNT TO THE IMPLANT. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED.

Additional Manufacturer Narrative · 0

ZIMMERBIOMET COMPLAINT NUMBER (B)(4). AGE AND DATE OF BIRTH UNKNOWN / NOT PROVIDED. WEIGHT UNKNOWN / NOT PROVIDED. PMA/510(K) NUMBERS: K011028, K013227. CUSTOMER HAS INDICATED THAT THE PRODUCT IS IN PROCESS OF BEING RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION WAS RECEIVED AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE SCREW BECAME STUCK INSIDE THE IMPLANT BODY ON TOOTH SITE #2 AND HAD TO BE REMOVED. ANOTHER IMPLANT WAS PLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
344960 IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 8MM DENTAL IMPLANT DZE ZIMMER DENTAL TSVWB8 1243473 00889024020085

Patients

Seq Age Sex Outcome Treatment
1 Female