FREESTYLE LIBRE 3
Report
- Report Number
- 2954323-2022-44112
- Event Type
- Injury
- Date Received
- December 2, 2022
- Date of Event
- November 4, 2022
- Report Date
- December 2, 2022
- Manufacturer
- ABBOTT DIABETES CARE LTD
- Product Code
- QLG
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- 003
Narratives
AT THIS TIME, PRODUCT HAS NOT YET BEEN RETURNED AND A VALID SERIAL NUMBER HAS NOT BEEN PROVIDED. AN EXTENDED INVESTIGATION HAS BEEN PERFORMED FOR THE REPORTED COMPLAINT AND IT HAS BEEN DETERMINED THAT THERE WAS NO INDICATION THAT THE PRODUCT DID NOT MEET SPECIFICATION. A TRIPPED TREND REVIEW WAS CONDUCTED FOR THE REPORTED COMPLAINT AND FS LIBRE READER, NO TRENDS WERE IDENTIFIED THAT WOULD INDICATE ANY PRODUCT RELATED ISSUES. IF THE PRODUCT IS RETURNED, A PHYSICAL INVESTIGATION WILL BE PERFORMED AND A FOLLOW-UP REPORT SUBMITTED. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, MODEL NUMBER 72111-01 WHICH HAS A SIMILAR PRODUCT DISTRIBUTED IN THE U.S., MODEL NUMBER 71953-01. THE DEVICE MANUFACTURING DATE IS UNKNOWN. THE DATE ENTERED IS THE DATE ABBOTT DIABETES CARE BECAME AWARE OF THE EVENT. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.
A SIGNAL LOSS ALARM ISSUE WAS REPORTED WITH THE ADC DEVICE. A CUSTOMER REPORTED THAT THE LOSS OF SIGNAL RESULTED IN THE HYPOGLYCEMIA. THE CUSTOMER EXPERIENCED SYMPTOMS OF DIZZINESS AND HAD NO STRENGTH. THE CUSTOMER HAD CONTACT WITH DOCTOR AND A READING OF 49 MG/DL WAS OBTAINED ON THE HCP METER AND WAS TREATED WITH AN UNSPECIFIED INJECTION. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1797460 | FREESTYLE LIBRE 3 | CONTINUOUS GLUCOSE MONITORING SYSTEM | QLG | ABBOTT DIABETES CARE LTD | 71953-01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Required Intervention |