ADVANCED CONTROL I-BASE, DC
Report
- Report Number
- 2921578-2022-00024
- Event Type
- Malfunction
- Date Received
- December 2, 2022
- Date of Event
- November 21, 2022
- Report Date
- February 24, 2023
- Manufacturer
- MIZUHO OSI
- Product Code
- JEA
- UDI-DI
- 00842430106026
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
BASED ON THE INFORMATION RECEIVED FROM THE INVESTIGATION, IT WAS DEEMED AS A USER ERROR AS THERE WAS MISUSE OF MIZUHO OSI TABLE ALONG WITH MAZOR X STEATH (NON-MIZUHO OSI PRODUCT) CONTRARY TO PUBLISHED INSTRUCTIONS BY MAZOR/MEDTRONIC THAT MENTIONED "THE SURGICAL TABLE MOVEMENT FEATURE MUST BE DISABLED DURING SURGERY WITH THE MAZOR X STEALTH EDITION SURGICAL SYSTEM". TABLE WAS ACTIVELY TILTED WITH THE MAZOR X ARM ATTACHED, WHICH CAUSED FAILURE OF THE TILT ACTUATOR DUE TO THE EXCESS MOVEMENT LOAD IMPARTED BY THE MAZOR X STEALTH ROBOT SYSTEM/ARM. IT IS ALSO BELIEVED THAT THERE WAS A MISUSE BY LACK OF REGULAR PREVENTATIVE MAINTENANCE OF A TABLE THAT IS 14+ YEARS OLD ALONG WITH OF DOCUMENTED YEARLY PREVENTATIVE MAINTENANCE. THE ADJUSTMENT OF THE HEAD AND FOOT END BRAKES WAS ALSO FOUND TO BE NON-COMPLIANT WITH THE SPECIFICATIONS MENTIONED IN THE OWNER'S MANUAL.
IT WAS REPORTED THAT MIZUHO OSI TABLE WAS BEING USED WITH A MAPPING SYSTEM AND STAFF UNLOCKED THE TABLE TO ROLL IT BUT PATIENT ROLLED OFF THE TABLE IN DOING SO. NO REPORTED PATIENT INJURY. HOSPITAL STAFF SLOWLY LOWERED THE PATIENT TO THE GROUND. THE PATIENT WAS LIFTED BACK TO ANOTHER BED FOR IMAGING AND SAFETY PRECAUTIONS. CASE WAS CANCELLED AND RESCHEDULED FOR NEXT DAY.
IT WAS REPORTED THAT MIZUHO OSI TABLE WAS BEING USED WITH A MAPPING SYSTEM AND STAFF UNLOCKED THE TABLE TO ROLL IT BUT PATIENT ROLLED OFF THE TABLE IN DOING SO.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2292341 | ADVANCED CONTROL I-BASE, DC | OPERATING TABLE | JEA | MIZUHO OSI | 5803 | 00842430106026 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |