FDA Adverse Event Malfunction Summary report: N

INRATIO

MDR report key: 1589778 · Received January 18, 2010

Report

Report Number
2027969-2010-00061
Event Type
Malfunction
Date Received
January 18, 2010
Date of Event
December 29, 2009
Report Date
January 18, 2010
Manufacturer
BIOSITE INCORPORATED
Product Code
GJS
PMA / PMN Number
021923
Removal / Correction Number
NI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: INRATIO: 4.9; REFERENCE: 2.6; MEAN: 3.75; CONFIDENCE LIMITS: 2.2-5.3. MEAN CALCULATION FROM COMPARISON TEST DATA PROVIDED BY END-USER REVEALED BOTH INRATIO AND LAB VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. AS OF 01/07/2010, THREE DISCREPANT RESULT COMPLAINTS WERE REPORTED FOR LOT #210832 YIELDING A COMPLAINT RATE OF 0.005%. DUE TO THIS LOW OCCURRENCE RATE, BELOW THE ACTION THRESHOLDS MONITORED BY QA FOR CORRECTIVE ACTION (>0.07%) NO FURTHER ACTION IS REQUIRED AT THIS TIME. ONGOING TRENDING SHALL BE PERFORMED TO DETERMINE IF FURTHER ACTION IS WARRANTED. NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME AS NO PRODUCT DEFICIENCY WAS ESTABLISHED.

Description of Event or Problem · 1

CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB. RESULTS AS FOLLOWS: DATE: (B)(6) 2009; INRATIO: 4.9; LAB: 2.6.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INRATIO PROTHROMBIN TIME TEST GJS BIOSITE INCORPORATED 100139 210832