FDA Adverse Event Injury Summary report: N

NC EMERGE

MDR report key: 15896249 · Received December 1, 2022

Report

Report Number
2124215-2022-50136
Event Type
Injury
Date Received
December 1, 2022
Date of Event
July 9, 2022
Report Date
December 1, 2022
Manufacturer
BOSTON SCIENTIFIC CORPORATION
Product Code
LOX
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MATSUKAWA, R., MATSUURA, H., TOKUTOME, M., OKAHARA, A., HARA, A., OKABE, K., KAWAI, S., & MUKAI, Y. (2022). USE OF A CUTTING BALLOON REDUCES THE INCIDENCE OF DISTAL EMBOLISM IN ACUTE CORONARY SYNDROME REQUIRING PREDILATATION BEFORE STENTING. CIRCULATION REPORTS, 4(8), 345-352. HTTPS://DOI.ORG/10.1253/CIRCREP.CR-22-0056.

Description of Event or Problem · 0

IT WAS REPORTED VIA JOURNAL ARTICLE THAT DISTAL EMBOLISM OCCURRED. THIS STUDY UTILIZED PSEUDO-THROMBUS IN AN EX VIVO EXPERIMENTAL MODEL TO COMPARE THROMBUS DISPERSAL USING A CUTTING BALLOON AND NON-COMPLIANT BALLOON. THERE WERE NO PATIENTS INVOLVED IN THIS MODEL. THE PSEUDO-THROMBUS LESIONS WERE DILATED WITH EITHER A CUTTING BALLOON (WOLVERINE 3.0X10MM, 6ATM) OR A NON-COMPLIANT BALLOON (EMERGE 3.0X12MM, AT 12ATM) FOR 10 SECONDS EACH. THE WEIGHT OF THE PSEUDO-THROMBUS MODEL WAS MEASURED BEFORE AND AFTER DILATATION. THE DIFFERENCE IN WEIGHT FROM BEFORE TO AFTER BALLOON DILATATION WAS DIVIDED BY THE ORIGINAL WEIGHT TO OBTAIN THE PSEUDO-THROMBUS DISPERSAL RATE. PSEUDO-THROMBUS DISPERSAL, REGARDED AS AN INDEX OF DISTAL EMBOLI, WAS SIGNIFICANTLY LOWER IN THE CUTTING BALLOON THAN THE NON-COMPLIANT BALLOON.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2769201 NC EMERGE CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other