FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 15891749 · Received December 1, 2022

Report

Report Number
9610877-2022-60578
Event Type
Malfunction
Date Received
December 1, 2022
Date of Event
November 9, 2022
Report Date
December 1, 2022
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
FDT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL ED34-I10T-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K163614. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE OBJECTIVE PRISM WAS CRACKED. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE OBJECTIVE PRISM. BASED ON THE TECHNICAL REPORT "HR-RPT-0586(IMAGE FAILURE)" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(OBJECTIVE LENS BROKEN).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1854044 PENTAX VIDEO DUODENOSCOPE - I10 STANDARD FDT HOYA CORPORATION PENTAX TOKYO OFFICE ED34-I10T

Patients

Seq Age Sex Outcome Treatment
1 Unknown