FDA Adverse Event Malfunction Summary report: N

ENDO GIA

MDR report key: 15885700 · Received November 30, 2022

Report

Report Number
2647580-2022-04350
Event Type
Malfunction
Date Received
November 30, 2022
Date of Event
November 8, 2022
Report Date
November 30, 2022
Manufacturer
US SURGICAL PUERTO RICO
Product Code
GDW
UDI-DI
10884523003215
PMA / PMN Number
K111825
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

CONCOMITANT MEDICAL PRODUCTS: SIG POWER SIGADAPTSTND LINEAR ADAPTER (SN:(B)(4)); EGIA60AMT EGIA 60 ARTIC MED THICK SULU (LOT#P2F0017); SIG POWER SIGPSHELL CONTROL SHELL (LOT#N2H0505Y); SIG POWER SIGPHANDLE HANDLE (LOT#C20AAF0230). EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS AVAILABLE FOR EVALUATION. THE EVALUATION FOUND NO POTENTIALLY CONTRIBUTING FACTORS, AND THE SAMPLE MET ALL RELATED SPECIFICATIONS. IT WAS REPORTED THAT THE JAWS OF THE RELOAD DID NOT CLOSE COMPLETELY. THE REPORTED ISSUE COULD NOT BE CONFIRMED. THE MOST LIKELY CAUSE COULD NOT BE IDENTIFIED BECAUSE NO RELATED PROBLEM WAS DETECTED WITH THE DEVICE. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC ASSISTED DISTAL GASTRECTOMY PROCEDURE, THE CARTRIDGE WAS CONNECTED AT THE TIME OF THE 4TH REMAINING GASTRODUODENAL ANASTOMOSIS. THERE WAS A CATCH WHEN THE PORT WAS INSERTED, WHEN IT WAS CHECKED, THE JAW DID NOT CLOSE NORMALLY. AFTERWARDS, IT WAS REPLACED WITH A NEW CARTRIDGE, BUT IN THE CARTRIDGE ALSO THE JAW DID NOT CLOSE COMPLETELY. THE ADAPTER HAS AN ERROR DISPLAY, POST-OPERATIVELY. A COMPETITOR DEVICE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1729395 ENDO GIA STAPLE, IMPLANTABLE GDW US SURGICAL PUERTO RICO EGIA60AMT P2F0017 10884523003215

Patients

Seq Age Sex Outcome Treatment
1 Male