FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP MMT-551NAH

MDR report key: 15883945 · Received November 30, 2022

Report

Report Number
2032227-2022-367391
Event Type
Malfunction
Date Received
November 30, 2022
Date of Event
November 27, 2022
Report Date
March 31, 2023
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
UDI-DI
000000643169503632
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 0

CASE TYPE = PARADIGM. ON 27-NOV-2022 CUSTOMER RETURNED PUMP FOR AN ALLEGED AA33 ALARM, LOOSE DRIVE SUPPORT DISK. UNIT RECEIVED MOTOR ERROR ALARM DURING THE BASIC OCCLUSION TEST DUE TO LOOSE/PROTRUDED DRIVE SUPPORT DISK. UNABLE TO PERFORM BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME/A33 TEST, EXCESSIVE NO DELIVERY TEST OR VERIFY A33 ALARM, PRIME/FILL ANOMALY DUE TO MOTOR ERROR ALARM. HOWEVER, UNIT PASSED REWIND TEST AND DISPLACEMENT TEST. PUMP HISTORY DOWNLOAD USING THDS WAS SUCCESSFUL. HOWEVER, A47 ALARMS FOUND ON THE HISTORY FILE WHICH CAUSED CORRUPTED DATA. UNABLE TO CONFIRM THE REPORTED EVENT DATE DUE TO THE CORRUPTED DATA. A47 ALARMS ARE DUE TO THE PUMP NOT HAVING POWER FOR A LONG PERIOD OF TIME. NO MOISTURE DAMAGE FOUND ON THE ELECTRONICS, MOTOR, BATTERY TUBE AND VIBRATOR ASSEMBLY NOTED. THE TEST P-CAP/RESERVOIR DOES LOCK INTO PLACE. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: BROKEN BATTERY TUBE THREADS, PARTIAL BROKEN RESERVOIR TUBE LIP, BROKEN BELT CLIP RAIL, STAINED ADDRESS/SERIAL NUMBER LABEL, STAINED END CAP STICKER, CRACKED RESERVOIR TUBE WINDOW, CRACKED CASE AT RESERVOIR TUBE WINDOW CORNER. UNABLE TO CONFIRM A33 ALARM DUE TO MOTOR ERROR ALARM. UNIT RECEIVED MOTOR ERROR ALARM DURING TESTING DUE TO LOOSE/PROTRUDED DRIVE SUPPORT DISK WAS CONFIRMED. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECT ACTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT, CUSTOMER REPORTED COMPROMISED FORCE SENSOR SYSTEM ALARM (A33 ALARM). NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. TROUBLESHOOTING WAS PERFORMED AND THERE IS NO INDICATION THAT, INSULIN PUMP WAS DROPPED OR BUMPED. CUSTOMER WILL DISCONTINUE USE OF THE DEVICE, HOWEVER, INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2330412 530G INSULIN PUMP MMT-551NAH ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-551NAH A4551NAHJ 000000643169503632

Patients

Seq Age Sex Outcome Treatment
1 48 YR Female