FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 AG SELF TEST 2CT

MDR report key: 15882628 · Received November 29, 2022

Report

Report Number
1221359-2022-10199
Event Type
Malfunction
Date Received
November 29, 2022
Date of Event
November 5, 2022
Report Date
January 31, 2023
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA210264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 200884 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160 / LOT 200884, TEST BASE PART NUMBER 195-430H / LOT 195845. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 200884 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. HOWEVER, A POSSIBLE ASSIGNABLE ROOT CAUSE IS SAMPLE INTERFERENCE OR CROSS CONTAMINATION. H3 OTHER TEXT : DEVICE DISCARDED, SINGLE USE DEVICE

Additional Manufacturer Narrative · 0

THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. DEVICE DISCARDED, SINGLE USE DEVICE.

Description of Event or Problem · 0

THE CUSTOMER REPORTED A FALSE NEGATIVE RESULT WITH THE BINAXNOW COVID-19 AG SELF-TEST PERFORMED ON (B)(6)2022 CONFIRMATION TESTING WITH (PLATFORM UNKNOWN) AT A HOSPITAL GENERATED A POSITIVE RESULT THE CUSTOMER STATED THE PATIENT WAS SYMPTOMATIC (COUGH AND RUNNY NOSE). NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED A FALSE NEGATIVE RESULT WITH THE BINAXNOW COVID-19 AG SELF-TEST PERFORMED ON (B)(6) 2022. CONFIRMATION TESTING WITH (PLATFORM UNKNOWN) AT A HOSPITAL GENERATED A POSITIVE RESULT. THE CUSTOMER STATED THE PATIENT WAS SYMPTOMATIC (COUGH AND RUNNY NOSE). NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2465312 BINAXNOW COVID-19 AG SELF TEST 2CT CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM. QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 200884 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 63 YR Male