WAMPOLE LABORATORIES STAPH LATEX KIT
Report
- Report Number
- 1215667-2010-00001
- Event Type
- Malfunction
- Date Received
- January 15, 2010
- Date of Event
- December 15, 2009
- Report Date
- January 15, 2010
- Manufacturer
- NERL DIAGNOSTICS, LLC
- Product Code
- JWX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
ADD'L LOT # 1110049. IN RESPONSE TO THE RECEIPT OF THESE COMPLAINTS, WE CONDUCTED AN INVESTIGATION. BOTH OF THESE LOTS WERE MANUFACTURED FROM THE SAME BULK LOT OF STAPH LATEX REAGENT. WE REPEATED QUALITY CONTROL RELEASE TESTING ON RETENTION KITS AS WELL AS RETURNED KITS. OUR TESTING INDICATED REDUCED REACTIVITY (OUT OF SPECIFICATION) OF THE LATEX REAGENT IN THE PRESENCE OF THE RECOMMENDED POSITIVE CONTROL. BASED ON THESE RESULTS, WE CONDUCTED A RISK ASSESSMENT. A DECISION WAS MADE TO RECALL BOTH LOTS OF PRODUCT WHICH WERE SHIPPED TO ONE DISTRIBUTOR. WE ARE WORKING WITH THAT DISTRIBUTOR TO ENSURE THAT ALL END USERS ARE NOTIFIED OF THE RECALL. A SAMPLE LETTER TO THE DISTRIBUTOR IS ATTACHED. IN ADDITION, TO ENSURE THAT THIS PRODUCT MALFUNCTION DOES NOT RECUR, THE FOLLOWING MEASURES WERE TAKEN. THE SOP FOR STOCK CULTURE MAINTENANCE (B)(4) HAS BEEN UPDATED TO INCLUDE THE ORGANISMS UTILIZED FOR THE QC BULK AND RELEASE TESTING FOR THE STAPH LATEX KIT. APPROPRIATE QC PERSONNEL HAVE BEEN TRAINED TO THE NEW REVISION (B)(4).
ON (B)(6) 2009, IT WAS CONFIRMED THAT THE STAPH LATEX REAGENT WAS NOT SUFFICIENTLY REACTIVE WITH THE RECOMMENDED POSITIVE CONTROL. THE STAPH LATEX KIT IS A LATEX AGGLUTINATION KIT IN WHICH THE REAGENT SHOULD AGGLUTINATE WHEN TESTED WITH A KNOWN POSITIVE CONTROL ORGANISM (B)(6). WHEN THE KITS IN QUESTION WERE TESTED WITH THE POSITIVE CONTROL, AS RECOMMENDED IN THE PACKAGE INSERT, WEAKENED AGGLUTINATION WAS OBSERVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | WAMPOLE LABORATORIES STAPH LATEX KIT | JWX | NERL DIAGNOSTICS, LLC | 0919119 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |