PEGASUS BL 22GA X 1.00IN PRN-CAP Y
Report
- Report Number
- 8041187-2022-00686
- Event Type
- Malfunction
- Date Received
- November 28, 2022
- Date of Event
- November 1, 2022
- Report Date
- January 9, 2023
- Manufacturer
- BECTON DICKINSON MEDICAL (SINGAPORE)
- Product Code
- FOZ
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
H6: INVESTIGATION SUMMARY: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 2082677. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. ALTHOUGH PHOTOS WERE SUBMITTED FOR EVALUATION, THE AFFECTED SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT.
IT WAS REPORTED WHILE USING PEGASUS BL 22GA X 1.00IN PRN-CAP Y LEAKAGE OCCURRED. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: WHEN THE HEAD NURSE OF THE EMERGENCY DEPARTMENT IMPLANTED THE INDWELLING NEEDLE TO THE PATIENT FOR INFUSION, SHE FOUND THAT THE EXTENSION TUBE CLIPPED BY THE INFUSION CLIP WAS LEAKING FLUID.
IT WAS REPORTED WHILE USING PEGASUS BL 22GA X 1.00IN PRN-CAP Y LEAKAGE OCCURRED. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: WHEN THE HEAD NURSE OF THE EMERGENCY DEPARTMENT IMPLANTED THE INDWELLING NEEDLE TO THE PATIENT FOR INFUSION, SHE FOUND THAT THE EXTENSION TUBE CLIPPED BY THE INFUSION CLIP WAS LEAKING FLUID.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2112824 | PEGASUS BL 22GA X 1.00IN PRN-CAP Y | INTRAVASCULAR CATHETER | FOZ | BECTON DICKINSON MEDICAL (SINGAPORE) | 2082677 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |