Report

Report Number
3013886523-2022-00553
Event Type
Injury
Date Received
November 28, 2022
Report Date
March 22, 2023
Manufacturer
INTEGRA LIFESCIENCES SWITZERLAND SAR
Product Code
JXG
UDI-DI
10381780519164
PMA / PMN Number
K053107
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION. UPON COMPLETION OF THE INVESTIGATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THE HAKIN VALVE WAS NOT RETURNED FOR EVALUATION AFTER THREE GOOD FAITH ATTEMPTS (GFES) WERE MADE. LOT NUMBER INFORMATION HAS BEEN PROVIDED; THEREFORE, DEVICE HISTORY RECORD (DHR) WAS REVIEWED, AND NO ANOMALIES WERE FOUND. THE ROOT CAUSE OF THE REPORTED ISSUE COULD NOT BE DETERMINED. HOWEVER, A PROBABLE ROOT CAUSE FOR THE REPORTED COMPLAINT IS DUE TO BIOLOGICAL DEBRIS AND PROTEIN BUILD UP INTERFERING WITH THE VALVE MECHANISM. IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE IN THE FUTURE, THIS COMPLAINT WILL BE REOPENED, AND THE RESPECTIVE EVALUATION PERFORMED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR ISSUES. AT PRESENT, WE CONSIDER THIS COMPLAINT TO BE CLOSED.

Description of Event or Problem · 0

A PHYSICIAN REPORTED A HAKIM VALVE (ID 823113) WAS IMPLANTED ON (B)(6) 2022. THE PATIENT CAME IN WITH SHUNT OVER DRAINAGE. MEDICAL STAFF NEEDED TO SET THE VALVE PRESSURE, HOWEVER, THEY WERE UNABLE TO REPROGRAM THE VALVE AFTER MULTIPLE ATTEMPTS AND A VALVE REVISION WAS REQUIRED. OPERATION PERFORMED; "ENDOSCOPIC SEPTOSTOMY AND LEFT FRONTAL KOCHER'S POINT PROGRAMMABLE VP SHUNT UNDER GA."

Description of Event or Problem · 0

N/A

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2713045 MICRO CHPV W RICKHAM UNITIZED CHPV JXG INTEGRA LIFESCIENCES SWITZERLAND SAR 823113 5346513 10381780519164

Patients

Seq Age Sex Outcome Treatment
1 Male