ORTHO SUMMIT SAMPLE HANDLING SYSTEM

MDR report key: 1586762 · Received January 27, 2010

Report

Report Number
2250051-2010-00021
Event Type
Malfunction
Date Received
January 27, 2010
Date of Event
January 18, 2010
Report Date
January 27, 2010
Manufacturer
HAMILTON BONADUZ AG
Product Code
JTC
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT HAS BEEN INVESTIGATED. THE FIELD SERVICE ENGINEER (FSE) FOUND A DEFECTIVE PLUNGER CLAMP IN THE REPORTED PROBLEM AREA OF THE PIPETTE ASSEMBLY. FSE REPLACED THE PLUNGER CLAMP IN POSITION # 3. FSE PERFORMED SPLLD CHECKING PROCEDURE AND TESTED SUMMIT WITH OAS USER SOFTWARE. THE INSTRUMENT WAS TESTED WITHOUT PROBLEM AND WAS RETURNED TO EXPECTED OPERATION.

Description of Event or Problem · 1

THE ORTHO SUMMIT SAMPLE HANDLING SYSTEM INSTRUMENT DID NOT PIPETTE SAMPLE AND DID NOT GENERATE AN ERROR MESSAGE. NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT.THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC. (B)(4)