FDA Adverse Event
Malfunction
Summary report: N
ORTHO SUMMIT SAMPLE HANDLING SYSTEM
MDR report key: 1586762
·
Received January 27, 2010
Report
- Report Number
- 2250051-2010-00021
- Event Type
- Malfunction
- Date Received
- January 27, 2010
- Date of Event
- January 18, 2010
- Report Date
- January 27, 2010
- Manufacturer
- HAMILTON BONADUZ AG
- Product Code
- JTC
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT HAS BEEN INVESTIGATED. THE FIELD SERVICE ENGINEER (FSE) FOUND A DEFECTIVE PLUNGER CLAMP IN THE REPORTED PROBLEM AREA OF THE PIPETTE ASSEMBLY. FSE REPLACED THE PLUNGER CLAMP IN POSITION # 3. FSE PERFORMED SPLLD CHECKING PROCEDURE AND TESTED SUMMIT WITH OAS USER SOFTWARE. THE INSTRUMENT WAS TESTED WITHOUT PROBLEM AND WAS RETURNED TO EXPECTED OPERATION.
Description of Event or Problem · 1
THE ORTHO SUMMIT SAMPLE HANDLING SYSTEM INSTRUMENT DID NOT PIPETTE SAMPLE AND DID NOT GENERATE AN ERROR MESSAGE. NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT.THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC. (B)(4)
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ORTHO SUMMIT SAMPLE HANDLING SYSTEM | SAMPLE PIPETTER | JTC | HAMILTON BONADUZ AG | 7003009 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |