FDA Adverse Event Other Summary report: N

OMNIPOD INSULIN PUMP

MDR report key: 1586133 · Received January 21, 2010

Report

Report Number
3004464228-2010-01030
Event Type
Other
Date Received
January 21, 2010
Date of Event
December 21, 2009
Report Date
December 22, 2009
Manufacturer
INSULET CORPORATION
Product Code
LZG
PMA / PMN Number
K042792
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT WILL NOT BE RETURNED TO THE MFR FOR EVALUATION. WE ARE UNABLE TO CONFIRM ANY PRODUCT MALFUNCTION. NO CONCLUSION CAN BE REACHED AT THIS TIME. THE CUSTOMER STATED THAT A KINK WAS NOTED IN THE CANNULA, THERE WAS, HOWEVER, NO REPORT OF AN OCCLUSION ALARM. THE POD WOULD HAVE INITIATED AN OCCLUSION ALARM IF THE FLOW OF INSULIN WAS RESTRICTED/PREVENTED BY THE KINK AND RESULTED IN BACK PRESSURE. THE PRODUCT USER GUIDE INSTRUCTS THE USER TO "CHECK INFUSION SITE FREQUENTLY FOR PROPER CANNULA PLACEMENT." IT ALSO SUGGESTS THAT THE USER SHOULD CHECK THEIR BLOOD GLUCOSE LEVELS FREQUENTLY IN ORDER TO NOTICE AND REACT QUICKLY AND APPROPRIATELY TO ANY ISSUES. THE PT REMEDIED THE SITUATION BY REMOVING AND REPLACING THE DEVICE PER USER INSTRUCTIONS.

Description of Event or Problem · 1

CUSTOMER REPORTED EXPERIENCING HIGH BGS (280-460MG/DL) WHILE WEARING A POD. CUSTOMER REPORTED THAT THERE WAS A KINK IN THE CANNULA BUT THE POD DID NOT INITIATE AN ALARM. PRODUCT WILL NOT BE RETURNED FOR INVESTIGATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 11200 NA

Patients

Seq Age Sex Outcome Treatment
1 11 YR Other