REVERSE SHOULDER SYSTEM HUMERAL REVERSE HC LINER Ø39/+0MM
Report
- Report Number
- 3005180920-2022-00902
- Event Type
- Injury
- Date Received
- November 24, 2022
- Date of Event
- November 7, 2022
- Report Date
- November 24, 2022
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- PHX
- UDI-DI
- 07630040706292
- PMA / PMN Number
- K170452
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- 003
Narratives
BATCH REVIEW PERFORMED ON (B)(6) 2022: LOT 2203309: (B)(4) ITEMS MANUFACTURED AND RELEASED ON (B)(6) 2022. EXPIRATION DATE: 2027-04-07. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED CASE DURING THE PERIOD OF REVIEW. ADDITIONAL DEVICES INVOLVED: BATCH REVIEW PERFORMED ON (B)(6) 2022: REVERSE SHOULDER SYSTEM 04.01.0211 LAT. GLENOSPHERE 39XØ27 (K193175) LOT 2212985: (B)(4) ITEMS MANUFACTURED AND RELEASED ON (B)(6) 2022. EXPIRATION DATE: 2027-07-19. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED CASE DURING THE PERIOD OF REVIEW. REVERSE SHOULDER SYSTEM 04.01.0182 SHORT HUMERAL DIAPHYSIS - CEMENTLESS - 9 (K180089) LOT 2202582: (B)(4) ITEMS MANUFACTURED AND RELEASED ON (B)(6) 2022. EXPIRATION DATE: 2027-03-22. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 25 ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED CASE DURING THE PERIOD OF REVIEW.
AT ABOUT 1 MONTH AFTER THE PRIMARY SURGERY, THE PATIENT CAME IN REPORTING PAIN DUE TO JOINT LUXATION. DURING THE SURGERY, IT WAS NOTICED THAT THE SHORT STEM WAS A LITTLE LOOSE DUE TO POOR BONE QUALITY. THE SURGEON REVISED SUCCESSFULLY THE STEM, METAPHYSIS, LINER AND GLENOSPHERE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1918865 | REVERSE SHOULDER SYSTEM HUMERAL REVERSE HC LINER Ø39/+0MM | SHOULDER HUMERAL LINER | PHX | MEDACTA INTERNATIONAL SA | 04.01.0122 | 2203309 | 07630040706292 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Male | Required Intervention |