FDA Adverse Event Injury Summary report: N

STEM: AMISTEM H HA COATED LAT STEM SIZE 4

MDR report key: 15860278 · Received November 24, 2022

Report

Report Number
3005180920-2022-00870
Event Type
Injury
Date Received
November 24, 2022
Date of Event
November 16, 2022
Report Date
February 8, 2023
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030804205
PMA / PMN Number
K093944
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

DEVICE TO BE RECEIVED AND ANALYZED. X-RAYS ANALYSIS TO BE PERFORMED. ADDITIONAL INFORMATION AND RESULTS OF THE ANALYSIS COULD BE AVAILABLE IN THIS CASE FOLLOW UP 1.

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 9 JANUARY 2023: LOT 112384: 10 ITEMS MANUFACTURED AND RELEASED ON 05-SEP-2011. EXPIRATION DATE: 2016-JUL-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ALL ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS DEPARTMENT REVISION APPROXIMATELY 6 YEARS AFTER INDEX SURGERY IN A MALE PATIENT DUE TO ASEPTIC LOOSENING. FROM THE IMAGES, SIGNS OF STRESS SHIELDING AND RADIOGRAPHIC LOOSENING ARE VISIBLE. ASEPTIC LOOSENING IS A POSSIBLE LITERATURE DESCRIBED ADVERSE EVENT AFTER PRIMARY CEMENTLESS HIP ARTHROPLASTIES AND ITS CAUSES ARE OFTEN UNKNOWN. THE REASON FOR THIS FAILURE CANNOT BE DETERMINED. VISUAL INSPECTION PERFORMED BY R&D PROJECT MANAGER DURING THE ANALYSIS IT WAS EVALUATED THAT ALMOST THE PROXIMAL HALF OF THE STEM BODY IS STILL COVERED WITH HA COATING. THIS IN UNEXPECTED CONSIDERING THAT THE STEM WAS IMPLANTED SIX YEARS AGO. SOME SIGNS AND SCRATCHES ARE PRESENT ON THE NECK SURFACE, PROBABLY DUE TO REVISION SURGERY. IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF REPORTED LOOSENING.

Description of Event or Problem · 0

AMISTEM IMPLANTED APPROXIMATELY 6 YEARS AGO. REVISION SURGERY PERFORMED DUE TO ASEPTIC LOOSENING. WAITING FOR INFO REGARDING REVISION.

Description of Event or Problem · 0

REVISION SURGERY PERFORMED APPROXIMATELY 10 YEARS AFTER THE PRIMARY SURGERY DUE TO ASEPTIC LOOSENING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1917902 STEM: AMISTEM H HA COATED LAT STEM SIZE 4 CEMENTLESS STEM LZO MEDACTA INTERNATIONAL SA 01.18.144 112384 07630030804205

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention