COMPRESSION/DISTRACTION DEVICE F/FOOT+AN
Report
- Report Number
- 8030965-2022-10201
- Event Type
- Malfunction
- Date Received
- November 23, 2022
- Date of Event
- January 1, 2022
- Manufacturer
- SYNTHES GMBH
- Product Code
- HWN
- UDI-DI
- 07611819375819
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. A REVIEW OF THE DEVICE HISTORY RECORDS WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER, AND AN NC WAS STARTED BECAUSE THE VISUAL INSPECTION IN THE FINAL INSPECTION FOUND THAT THE SUBCOMPONENT 208.0421 (SWING ARM) WERE DAMAGED ON THE THREAD (M6XPH1.5 P0.75). THE DEVICES WERE VISUAL INSPECTED TO 100%, 10 DEVICES WERE DAMAGED ON THE THREAD AND WERE SCRAPPED. THE NC HAVE NO IMPACT ON THE COMPLAINT ISSUE. THE RAW MATERIAL CERTIFICATE WAS REVIEWED AND THE USED MATERIAL WAS ACCORDING TO THE SPECIFICATION OF THE DEVICE. PART NUMBER: 03.111.021, LOT NUMBER: T160928, MANUFACTURING SITE: TUTTLINGEN, RELEASE TO WAREHOUSE DATE: 22-NOV-2018. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THIS MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION BUT HAS YET TO BE RECEIVED. REPORTER IS A J&J SALES REPRESENTATIVE. (B)(4). THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H10 ADDITIONAL NARRATIVE: D9: DEVICE RETURNED. H6: THE PRODUCT WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. THE DEPUY SYNTHES TEAM CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THERE WAS NO DAMAGE OR DEFECTS WITH THE COMPRESSION/DISTRACTION DEVICE F/FOOT+AN, P/N: 03.111.021. A DIMENSIONAL INSPECTION WAS NOT PERFORMED AS IT IS NOT APPLICABLE TO THE COMPLAINT CONDITION. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THE OVERALL COMPLAINT WAS NOT CONFIRMED AS THE COMPRESSION/DISTRACTION DEVICE F/FOOT+AN, P/N: 03.111.021 WAS FOUND TO HAVE NO DAMAGE OR DEFECTS. NO DEFINITIVE ROOT CAUSE COULD BE DETERMINED. THERE WAS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE CONTRIBUTED TO THE COMPLAINT. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DRAWING/SPECIFICATIONS REVIEWED? YES, REVIEWED. DIMENSIONAL INSPECTION: N/A. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. H6 TYPE OF INVESTIGATION: ANALYSIS OF PRODUCTION RECORDS WAS REPORTED ON PREVIOUS REPORT.
IT WAS REPORTED ON AN UNKNOWN DATE THAT THESE ARE BROKEN OR UNUSABLE PRODUCTS. . IT IS UNKNOWN WHEN THE ISSUE WAS DISCOVERED. THERE WAS NO PATIENT OUTCOME/CONSEQUENCE. THIS REPORT IS FOR ONE (1)COMPRESSION/DISTRACTION DEVICE F/FOOT+AN THIS IS REPORT 5 OF 6 FOR COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 362314 | COMPRESSION/DISTRACTION DEVICE F/FOOT+AN | INSTRUMENT, COMPRESSION | HWN | SYNTHES GMBH | T160928 | 07611819375819 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | 6.0MM HAND REAMER 450MM.| DRIVING CAP/THREADED.| DRIVING CAP/THREADED.| RADIOLUCENT INSERTION HANDLE FRN.| RADIOLUCENT INSERTION HANDLE FRN.| SHARP HOOK.| SM FRAG LCP INST & IMPL SET GRAPHIC CASE. |