FDA Adverse Event Injury Summary report: N

INCLUSIVE TAPERED IMPLANT 3.2 MMD X 10 MML X 3.0 MMP

MDR report key: 15845196 · Received November 22, 2022

Report

Report Number
3011649314-2022-00651
Event Type
Injury
Date Received
November 22, 2022
Date of Event
August 1, 2022
Report Date
June 19, 2024
Manufacturer
PRISMATIK DENTALCRAFT, INC.
Product Code
DZE
PMA / PMN Number
K121406
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE HAS NOT BEEN RETURNED. IF/ WHEN THE DEVICE IS RETURNED AN INVESTIGATION WILL BE CARRIED OUT AND A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS IS THE 2ND OF 3 FAILED IMPLANTS REPORTED ON THE SAME PATIENT. REFERENCE THE FOLLOWING MANUFACTURER REPORTS FOR THE REMAINING IMPLANTS: 3011649314-2022-00650 AND 3011649314-2022-00652.

Additional Manufacturer Narrative · 0

H6: MEDICAL DEVICE PROBLEM CODE WAS UPDATED TO REFLECT LOSS OF INTEGRATION, AS IT WAS NOTED INCORRECTLY ON THE INITIAL AND SUPPLEMENTAL SUBMISSIONS.  THIS COMPLAINT WILL BE KEPT ON RECORD FOR TRACK AND TRENDING PURPOSES.

Additional Manufacturer Narrative · 0

CAPA 23-00. MANUFACTURER REFERENCE: (B)(4).

Additional Manufacturer Narrative · 0

THE DEVICE INVESTIGATION HAS BEEN COMPLETED AND THE RESULTS ARE AS FOLLOWS: DHR RESULTS: THE DHR WAS REVIEWED FOR LOT# 6062234 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE PART MET ALL THE CRITERIA CALLED FOR IN THE PRODUCTION ROUTER. STOCK PRODUCT REVIEWED RESULTS: THE PACKAGED STOCK PRODUCT IS NOT APPLICABLE FOR REVIEW SINCE NO DEFECT WAS OBSERVED FROM THE RETURNED PRODUCT. INVESTIGATION METHODS/RESULTS: THE DEVICE WAS RETURNED BUT NOT IN ORIGINAL PACKAGE. THE DEVICE WAS VERIFIED TO BE AN INCLUSIVE TAPERED IMPLANT 3.2 MMD X 10 MML X 3.0 MMP (70-1070-IMP0036) USING RADIOGRAPHIC TEMPLATE (GD-437-091015). THERE WAS NO DEFECT OR NON-CONFORMITY OBSERVED AND THE THREADS WERE INTACT. MINIMAL BONE DEBRIS WAS OBSERVED ON THE IMPLANT. ROOT CAUSE: "LOSS OF OSSEOINTEGRATION" IS A COMMON COMPLAINT IN REGARDS TO IMPLANT FAILURE. THIS OCCURS WHEN THE PATIENT'S BONE DOES NOT INTEGRATE WITH THE IMPLANT SURFACE. THE POSSIBLE RESPONSES TO THIS COMPLAINT COULD BE ATTRIBUTED TO VARIOUS CAUSES. ALTHOUGH THE ROOT CAUSE FOR LOSS OF OSSEOINTEGRATION IS INCONCLUSIVE AND SPECIFIC TO EACH CASE, PROBABLE CAUSES COULD BE THE LOSS OF PRIMARY STABILITY AT THE OSTEOTOMY SITE DUE TO INSUFFICIENT BONE OR POOR BONE QUALITY; EITHER THE BONE WAS TOO SOFT OR THE OPERATOR ERRED IN CREATING AN OSTEOTOMY BIGGER THAN THE SIZE OF THE IMPLANT DIAMETER. PREMATURE LOADING, PATIENT'S HEALTH, PERI-IMPLANTITIS, SMOKING, AND LACK OF ORAL HYGIENE MAY ALSO BE CONTRIBUTING FACTORS. IFU 3023579 REV 3.0 (INCLUSIVE DENTAL IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: "ABSOLUTE SUCCESS CANNOT BE GUARANTEED. FACTORS SUCH AS INFECTION, DISEASE AND INADEQUATE BONE QUALITY AND/OR QUANTITY CAN RESULT IN OSSEOINTEGRATION FAILURES FOLLOWING SURGERY OR INITIAL OSSEOINTEGRATION." IN ADDITION. IFU 3023579 REV 3.0 (INCLUSIVE DENTAL IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: THE IMPLANT SITE SHOULD BE INSPECTED FOR ADEQUATE BONE BY RADIOGRAPHS, PALPATIONS AND VISUAL EXAMINATION. DETERMINE THE LOCATION OF NERVES AND OTHER VITAL STRUCTURES AND THEIR PROXIMITY TO THE IMPLANT SITE BEFORE ANY DRILLING TO AVOID POTENTIAL INJURY, SUCH AS PERMANENT NUMBNESS TO THE LOWER LIP AND CHIN. ALSO, IFU 3023579 REV 3.0 (INCLUSIVE DENTAL IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN PRECAUTION SECTION: "MINIMIZING TISSUE DAMAGE IS CRUCIAL TO SUCCESSFUL IMPLANT OSSEOINTEGRATION. IN PARTICULAR, CARE SHOULD BE TAKEN TO ELIMINATE SOURCES OF INFECTION, CONTAMINANTS, SURGICAL AND THERMAL TRAUMA. RISK OF OSSEOINTEGRATION FAILURE INCREASES AS TISSUE TRAUMA INCREASES. ALL DRILLING PROCEDURES SHOULD BE PERFORMED AT 2000 RPM OR LESS UNDER CONTINUAL AND COPIOUS IRRIGATION. ALL SURGICAL INSTRUMENTS USED MUST BE IN GOOD CONDITION AND SHOULD BE USED CAREFULLY TO AVOID DAMAGE TO IMPLANTS OR OTHER COMPONENTS. IMPLANTS SHOULD BE PLACED WITH SUFFICIENT STABILITY; HOWEVER, EXCESSIVE INSERTION TORQUE MAY RESULT IN IMPLANT FRACTURE, OR FRACTURE OR NECROSIS OF THE IMPLANT SITE. THE PROPER SURGICAL PROTOCOL SHOULD BE STRICTLY ADHERED TO. THIS IS THE 2ND OF 3 FAILED IMPLANTS REPORTED ON THE SAME PATIENT. REFERENCE THE FOLLOWING MANUFACTURER REPORTS FOR THE REMAINING IMPLANTS. 3011649314-2023-00650 3011649314-2023-00652 THIS COMPLAINT WILL BE KEPT ON RECORD FOR TRACK AND TRENDING PURPOSES.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE INCLUSIVE TAPERED IMPLANT FAILED. THE PATIENT'S BONE TYPE IS III AND HAS A HISTORY OF SMOKING A PACK DAILY. THE PATIENT PRESENTED ON (B)(6) 2021 FOR A PRIMARY PROCEDURE ON TOOTH #6. THE PATIENT RETURNED ON (B)(6) 2022 WITHOUT COMPLAINT. UPON EXAMINATION, THE PROVIDER NOTED INFECTION. THE PROVIDER DETERMINED THAT THE IMPLANT LOST INTEGRATION, AND THE DEVICE WAS REMOVED. THE PATIENT'S STATUS IS LISTED AS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1918564 INCLUSIVE TAPERED IMPLANT 3.2 MMD X 10 MML X 3.0 MMP INCLUSIVE TAPERED IMPLANT DZE PRISMATIK DENTALCRAFT, INC. 70-1070-IMP0036 6062234

Patients

Seq Age Sex Outcome Treatment
1 70 YR Female Required Intervention