FDA Adverse Event Malfunction Summary report: N

MEDICHOICE DELUXE MODEL 50 CHROME EXAM LAMP

MDR report key: 1584294 · Received January 7, 2010

Report

Report Number
1584294
Event Type
Malfunction
Date Received
January 7, 2010
Report Date
January 7, 2010
Manufacturer
GOODWIN MANUFACTURING, INC.
Product Code
KZF
Product Problem
Yes
Report Source
User Facility report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE FACILITY BOUGHT A NEW GOOSENECK MEDI CHOICE EXAM LAMP FROM OWENS & MINOR. THE LABELING ON THE BOX INDICATES O & M GOT IT FROM GOODWIN MANUFACTURING. IT'S AN IMPORT & FOREIGN-MADE. LEAKAGE TESTS ON INCOMING SEEMED PECULIAR. THE HG (HOSPITAL GRADE) PLUG WAS OPENED AND THEY FOUND THE INSULATION CUT TO THE COPPER OF ALL THREE CONDUCTORS WHERE THE OUTER SHEATH WAS REMOVED. THE BLACK WIRE HAD ONLY THE BITTER END OF A SINGLE STRAND CLAMPED AND THE POLARITY WAS REVERSED. GOODWIN WAS CALLED AND ADVISED ABOUT THE PROBLEM BY THE FACILITY. THE FACILITY ADVISED OTHER STAFF TO BE ON WATCH FOR ANY BOX SAYING "DELUXE MODEL 50 CHROME EXAM LAMP."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MEDICHOICE DELUXE MODEL 50 CHROME EXAM LAMP GOOSENECK EXAM LAMP KZF GOODWIN MANUFACTURING, INC. * *

Patients

Seq Age Sex Outcome Treatment
1 *