MIC SAFETY PEG KIT WITH ENFIT CONNECTORS - 20 FR - PULL
Report
- Report Number
- 3006646024-2022-00024
- Event Type
- Injury
- Date Received
- November 19, 2022
- Report Date
- December 28, 2022
- Manufacturer
- AVANOS MEDICAL INC.
- Product Code
- KNT
- UDI-DI
- 00350770445991
- PMA / PMN Number
- K924065
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
THE ACTUAL COMPLAINT PRODUCT WAS NOT RETURNED FOR EVALUATION. A REVIEW OF THE DEVICE HISTORY RECORD IS IN-PROGRESS. ALL INFORMATION REASONABLY KNOWN AS OF 18-NOV-2022 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AVANOS MEDICAL, INC. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AVANOS MEDICAL, INC. AVANOS MEDICAL, INC. HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AVANOS MEDICAL, INC. COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN AVANOS MEDICAL, INC. PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY.
THE DEVICE HISTORY RECORD FOR THE REPORTED LOT NUMBER, 30201596, WAS REVIEWED AND DOCUMENTED THAT THE LOT WAS MANUFACTURED ACCORDING TO THE APPROVED MANUFACTURING PROCEDURES, SPECIFICATIONS, AND THEN RELEASED BY QUALITY ASSURANCE. ALL INFORMATION REASONABLY KNOWN AS OF (B)(6) 2022 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AVANOS MEDICAL, INC. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AVANOS MEDICAL, INC. AVANOS MEDICAL, INC. HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AVANOS MEDICAL, INC. COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO (B)(4), IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN AVANOS MEDICAL, INC. PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY.
AVANOS MEDICAL RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING TWO DIFFERENT EVENTS. THIS IS THE SECOND OF TWO REPORTS. REFER TO 2026095-2022-00127 FOR THE FIRST EVENT. IT WAS REPORTED THAT UPON PASSING PEG TUBE DOWN THROUGH ESOPHAGUS, THE BUMPER CAUSED A TEAR IN THE PATIENT'S ESOPHAGUS. THE PATIENT WAS REPORTED TO BE OK AT THE TIME OF THE CALL AND WAS DISCHARGED HOME WITH THE TUBE IN PLACE. MEDICAL TREATMENT REQUIRED IS UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2398545 | MIC SAFETY PEG KIT WITH ENFIT CONNECTORS - 20 FR - PULL | DH EF PEG INITIAL PLACEMENT PRODUCTS | KNT | AVANOS MEDICAL INC. | 8180-20 | 30201596 | 00350770445991 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Male | Other |