BD INTIMA-II CLOSED IV CATHETER SYSTEM
Report
- Report Number
- 3014704491-2022-00590
- Event Type
- Malfunction
- Date Received
- November 18, 2022
- Date of Event
- October 16, 2022
- Report Date
- November 28, 2022
- Manufacturer
- BD (SUZHOU)
- Product Code
- FOZ
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H6: INVESTIGATION SUMMARY: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 2082643. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, VISUAL ANALYSIS WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RETAINED DEVICES WERE FOUND TO BE FREE OF ANY DAMAGE OR OTHER ABNORMALITIES. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT. H3 OTHER TEXT : SEE H10.
INVESTIGATION SUMMARY IN RESPONSE TO THE EVENT REPORTED A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 2082643. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, VISUAL ANALYSIS WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RETAINED DEVICES WERE FOUND TO BE FREE OF ANY DAMAGE OR OTHER ABNORMALITIES. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT.
IT WAS REPORTED WHILE USING BD INTIMA-II CLOSED IV CATHETER SYSTEM THE TUBING DIFFICULT TO CLAMP. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: ON OCTOBER 16, 2022, WHEN THE NURSE WAS PREPARING FOR INFUSION FOR THE PATIENT, SHE FOUND THAT THE INDWELLING NEEDLE HAD NO BUCKLE AND WAS INCONVENIENT TO FIX..
IT WAS REPORTED WHILE USING BD INTIMA-II CLOSED IV CATHETER SYSTEM THE TUBING DIFFICULT TO CLAMP. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: ON (B)(6) 2022, WHEN THE NURSE WAS PREPARING FOR INFUSION FOR THE PATIENT, SHE FOUND THAT THE INDWELLING NEEDLE HAD NO BUCKLE AND WAS INCONVENIENT TO FIX.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 230243 | BD INTIMA-II CLOSED IV CATHETER SYSTEM | INTRAVASCULAR CATHETER | FOZ | BD (SUZHOU) | 2082643 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |