HAHN TAPERED IMPLANT 5.0 X 8MM
Report
- Report Number
- 3011649314-2022-00677
- Event Type
- Injury
- Date Received
- November 18, 2022
- Date of Event
- August 15, 2022
- Report Date
- June 14, 2024
- Manufacturer
- PRISMATIK DENTALCRAFT, INC.
- Product Code
- DZE
- PMA / PMN Number
- K143353
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
THE DEVICE HAS NOT BEEN RETURNED. IF/WHEN THERE IS MORE INFORMATION PROVIDED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE PATIENTS RACE WAS NOT PROVIDED WHEN ASKED.
THE DEVICE WAS RETURNED, THE INVESTIGATION HAS BEEN COMPLETED AND THE RESULTS ARE AS FOLLOWS: DHR RESULTS: THE DHR WAS REVIEWED FOR LOT #6093575 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE PART MET ALL THE CRITERIA CALLED FOR IN THE PRODUCTION ROUTER. STOCK PRODUCT REVIEWED RESULTS: THE PACKAGED STOCK PRODUCT IS NOT APPLICABLE FOR REVIEW SINCE NO DEFECT WAS OBSERVED FROM THE RETURNED PRODUCT. INVESTIGATION METHODS/RESULTS: THE DEVICE WAS RETURNED BUT NOT IN ORIGINAL PACKAGE. AN ABUTMENT WAS STILL ENGAGED ON THE DEVICE. THE DEVICE WAS VERIFIED TO BE A HAHN TAPERED IMPLANT Ø5.0 X 8 MM (70-1154-IMP0014) USING RADIOGRAPHIC TEMPLATE (PK-209-062515). THERE WAS NO DEFECT OR NON-CONFORMITY OBSERVED AND THE THREADS WERE INTACT. MINIMAL BONE DEBRIS WAS OBSERVED ON THE IMPLANT. THE COMPLAINT IS VERIFIED BASED ON THE RETURNED PART(S) BUT CANNOT CONFIRM THE FAILURE MODE. THERE WAS NO EVIDENCE FOUND THAT INDICATED THAT THE REPORTED ISSUE WAS CAUSED BY THE DEVICE ITSELF. ROOT CAUSE "LACK OF PRIMARY STABILITY" IS A COMMON COMPLAINT IN REGARDS TO IMPLANT FAILURE. THIS OCCURS WHEN THE PATIENT'S BONE DOES NOT INTEGRATE WITH THE IMPLANT SURFACE. THE POSSIBLE RESPONSES TO THIS COMPLAINT COULD BE ATTRIBUTED TO VARIOUS CAUSES. ALTHOUGH THE ROOT CAUSE FOR LACK OF PRIMARY STABILITY IS INCONCLUSIVE AND SPECIFIC TO EACH CASE, PROBABLE CAUSES COULD BE DUE TO INSUFFICIENT BONE OR POOR BONE QUALITY; EITHER THE BONE WAS TOO SOFT OR THE OPERATOR ERRED IN CREATING AN OSTEOTOMY BIGGER THAN THE SIZE OF THE IMPLANT DIAMETER. PREMATURE LOADING, PATIENT'S HEALTH, PERI-IMPLANTITIS, SMOKING, AND LACK OF ORAL HYGIENE MAY ALSO BE CONTRIBUTING FACTORS. IFU 570 REV 3.0 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN PRECAUTION SECTION SURGICAL PROCEDURES: "MINIMIZING TISSUE DAMAGE IS CRUCIAL TO SUCCESSFUL IMPLANT OSSEOINTEGRATION. IN PARTICULAR, CARE SHOULD BE TAKEN TO ELIMINATE SOURCES OF INFECTION, CONTAMINANTS, SURGICAL AND THERMAL TRAUMA. RISK OF OSSEOINTEGRATION FAILURE INCREASES AS TISSUE TRAUMA INCREASES. ALL DRILLING PROCEDURES SHOULD BE PERFORMED AT 2000 RPM OR LESS UNDER CONTINUAL AND COPIOUS IRRIGATION. ALL SURGICAL INSTRUMENTS USED MUST BE IN GOOD CONDITION AND SHOULD BE USED CAREFULLY TO AVOID DAMAGE TO IMPLANTS OR OTHER COMPONENTS. IMPLANTS SHOULD BE PLACED WITH SUFFICIENT STABILITY; HOWEVER, EXCESSIVE INSERTION TORQUE MAY RESULT IN IMPLANT FRACTURE, OR FRACTURE OR NECROSIS OF THE IMPLANT SITE. THE PROPER SURGICAL PROTOCOL SHOULD BE STRICTLY ADHERED TO." IFU 570 REV 3.0 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: "ABSOLUTE SUCCESS CANNOT BE GUARANTEED. FACTORS SUCH AS INFECTION, DISEASE AND INADEQUATE BONE QUALITY AND/OR QUANTITY CAN RESULT IN OSSEOINTEGRATION FAILURES FOLLOWING SURGERY OR INITIAL OSSEOINTEGRATION." IFU 570 REV 3.0 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: "THE IMPLANT SITE SHOULD BE INSPECTED FOR ADEQUATE BONE BY RADIOGRAPHS, PALPATIONS AND VISUAL EXAMINATION. DETERMINE THE LOCATION OF NERVES AND OTHER VITAL STRUCTURES AND THEIR PROXIMITY TO THE IMPLANT SITE BEFORE ANY DRILLING TO AVOID POTENTIAL INJURY, SUCH AS PERMANENT NUMBNESS TO THE LOWER LIP AND CHIN." THIS COMPLAINT WILL BE KEPT ON RECORD FOR TRACK AND TRENDING PURPOSES.
ADDITIONAL DATA: D4 (LOT NUMBER, EXPIRATION DATE, AND UDI NUMBER), H6 (HEALTH EFFECT - CLINICAL CODE, HEALTH EFFECT - IMPACT CODE - 4624, TYPE OF INVESTIGATION CODE, AND INVESTIGATION CONCLUSIONS CODE). CORRECTED DATA: B5, D4 (MODEL NUMBER AND CATALOG NUMBER), D6A, D6B, E1, H6 (HEALTH EFFECT - IMPACT CODE - 4627, AND MEDICAL DEVICE PROBLEM CODE). CAPA 23-006. MANUFACTURER REFERENCE: (B)(4).
IT WAS REPORTED THAT THE HAHN TAPERED IMPLANT FAILED. THE BONE GRADE IS NOT NOTED. THE PATIENT'S ORAL HYGIENE IS NOTED AS GOOD WITH A DENTAL HISTORY OF PREVIOUS MULTIPLE IMPLANTS AND CROWNS. THE PATIENT PRESENTED ON (B)(6)2022 FOR IMPLANT PLACEMENT ON TOOTH #18. AFTER THE HEALING CAP WAS PLACED THE PATIENT RETURNED (B)(6) 2022 WITH COMPLAINTS OF A LOOSE IMPLANT. UPON EXAM, THE PROVIDER NOTES THAT THE IMPLANT WAS LOOSE, SWELLING AND PUSS AROUND THE IMPLANT.
IT WAS REPORTED THAT THE HAHN TAPERED IMPLANT FAILED. THE BONE GRADE IS NOT NOTED. THE PATIENT'S ORAL HYGIENE IS NOTED AS GOOD. THE PATIENT PRESENTED ON (B)(6) 2022 FOR IMPLANT PLACEMENT ON TOOTH # 18. AFTER THE HEALING CAP WAS PLACED (1.5 MONTHS AFTER THE SURGERY), THE PATIENT RETURNED (B)(6) 2022 WITH COMPLAINTS OF A LOOSE IMPLANT. UPON EXAMINATION, THE PROVIDER NOTED LOOSE IMPLANT, INFECTION, PUS AND TISSUE SWELLING AROUND THE IMPLANT, AND THAT THE IMPLANT FAILED TO INTEGRATE. THE DEVICE WAS REMOVED AT THAT TIME AND BONE GRAFTING AND SUTURES WERE PLACED. THE AREA WILL BE LEFT TO HEAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 238934 | HAHN TAPERED IMPLANT 5.0 X 8MM | HAHN TAPERED IMPLANT | DZE | PRISMATIK DENTALCRAFT, INC. | 70-1154-IMP0014 | 6093575 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Female | Required Intervention |