FDA Adverse Event Injury Summary report: N

CYPHER SIROLIMUS-ELUTING CORONARY STENT

MDR report key: 1582795 · Received January 18, 2010

Report

Report Number
9616099-2010-00061
Event Type
Injury
Date Received
January 18, 2010
Date of Event
December 21, 2009
Report Date
December 21, 2009
Manufacturer
CORDIS DE MEXICO
Product Code
NIQ
PMA / PMN Number
NA
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TO REMOVE THE REMAINING PART OF THE SDS, A GC (BRITE TIP 7F XB3.0) WAS RE-ENGAGED AND THE SNARE CATHETER WAS DELIVERED TO THE LAD. THEN AN ANGIOGRAPHIC CATHETER (4F PIG TAIL) APPROACHED FROM THE FEMORAL ARTERY, WAS ALSO DELIVERED TO THE LAD. THE REMAINING PART OF THE SDS WAS PICKED UP WITH THE DISTAL CIRCLE OF THE ANGIOGRAPHIC CATHETER (4F PIG TAIL) AND THEN THE HUB OF THE ANGIOGRAPHIC CATHETER WAS SEVERED AND THE REMAINING PART OF THE SDS AND THE GC WERE SAFELY REMOVED FROM THE PT USING A SNARE FROM RADIAL ARTERY. TO FINISH THE PROCEDURE, A GW (X-TREME) RE-CROSSED THE LESION AND POST-DILATION WAS CONDUCTED TO FURTHER DILATE THE CYPHER STENT. AFTERWARDS, A SECOND CYPHER (3.5/13MM) WAS IMPLANTED PROXIMAL TO THE CYPHER, OVERLAPPING IT. THE PROCEDURE WAS FINISHED SUCCESSFULLY. THERE WAS NO PT INJURY REPORTED. THE PHYSICIAN COMMENTED THAT WHEN THE PROXIMAL PORTION OF THE STENT WAS INFLATED OUTSIDE THE PT, AN UNEXPECTED PORTION OF THE STENT WAS ALSO INFLATED, AND SO THE PORTION WAS SWAGED AND TWISTED TO MOUNT THE BALLOON. THE POSSIBLE CAUSE OF THE INFLATION DIFFICULTY WAS THAT THE STENT STRUTS MIGHT BECOME STUCK EACH OTHER OR THE BALLOON MIGHT BECOME STUCK WITH THE STENT STRUT WHEN THE STENT WAS SWAGED AND TWISTED. THE PRODUCT HAS BEEN RETURNED FOR EVALUATION AND TESTING; HOWEVER, THE ENGINEERING EVALUATION HAS NOT BEEN COMPLETED. ADDITIONAL INFO WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.

Description of Event or Problem · 1

DURING A CORONARY STENT PLACEMENT PROCEDURE, THE PHYSICIAN HAD DIFFICULTY PLACING THE STENT IN THE INTENDED TARGET LESION DUE TO INFLATION DIFFICULTIES WITH THE BALLOON. WHEN THE PHYSICIAN ATTEMPTED TO WITHDRAW THE STENT DELIVERY SYSTEM (SDS) FROM THE VESSEL, THE SDS SEPARATED AND NEEDED TO BE REMOVED WITH A SNARE. THE PT'S AGE WAS UNK, BUT WAS A FEMALE. THE TARGET LESION WAS THE PROXIMAL LEFT ANTERIOR DESCENDING (LAD) ARTERY. THE LESION WAS A DE NOVO, BIFURCATED, NOT CALCIFIED AND NOT TORTUOUS LESION. THERE WAS 90% STENOSIS. A RIGHT RADIAL APPROACH WAS CHOSEN FOR THE PROCEDURE. A BRITE TIP 6F XB3.0 GUIDE CATHETER WAS ENGAGED AND THEN A GUIDEWIRE (GW)(FIELDER) CROSSED THE TARGET LESION, AND ANOTHER GW (RINATO) WAS DELIVERED TO THE LEFT CIRCUMFLEX (LCX). PRE-DILATION WAS PERFORMED WITH A MAVERICK 2.5/15MM BALLOON CATHETER AND IVUS WAS CONDUCTED. THEN, A CYPHER (3.5/23MM: COMPLAINT PRODUCT) STENT, IN WHICH THE PROXIMAL PORTION OF THE STENT WAS INFLATED AT 5ATM OUTSIDE THE PT, AND A STRUT OF THE PROXIMAL PORTION OF THE STENT WAS CROSSED WITH GW BEFORE THE DELIVERY WAS DELIVERED TO THE TARGET LESION. WHEN THE CYPHER WAS INFLATED UP TO 25ATM, THE PROXIMAL AND DISTAL PORTION OF THE BALLOON WERE INFLATED BUT THE MID-PORTION OF THE BALLOON WOULD NOT BE INFLATED AT ALL. THEREFORE, THE PHYSICIAN TRIED TO DISPLACE THE BALLOON, BUT THE SDS OF THE CYPHER BECAME STUCK. WHEN THE PHYSICIAN PULLED STRONGLY THE SDS OF THE CYPHER, THE GUIDE WIRE EXIT-PORT AREA OF THE SHAFT WAS SEPARATED. THE TWO GWS (FIELDER AND RINATO) CAME OFF OF THE LEFT ANTERIOR DESCENDING (LAD) AND LCX AT THE TIME, SO THE DISTAL PART OF THE SDS REMAINED IN THE LAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CYPHER SIROLIMUS-ELUTING CORONARY STENT DRUG-ELUTING STENT (NIQ) NIQ CORDIS DE MEXICO NA 15041810

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L| R APEX 1.5/12MM| QUANTUM 3.0/8MM| STENT: CYPHER 3.5/13MM| X-TREME| BC: QUANTUM 3.5/15MM| MAVERICK 2.5/15MM| FIELDER FC| FIELDER| GW: RINATO| INFLATION DEVICE: NEICH'S| GC: BRITE TIP 6F XB3.0| BRITE TIP 7F XB3.0| SHEATH: TERUMO'S