BLUNT TIP SCREW, 4X48MM
Report
- Report Number
- 0009613350-2022-00608
- Event Type
- Injury
- Date Received
- November 18, 2022
- Date of Event
- October 20, 2022
- Report Date
- December 16, 2022
- Manufacturer
- ZIMMER MANUFACTURING SWITZERLAND GMBH
- Product Code
- HSB
- UDI-DI
- 00889024505483
- PMA / PMN Number
- K200814
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
(B)(4). MEDICAL DEVICES: PROXIMAL HUMERUS, RIGHT, 11X160MM, ITEM# 47249616011, LOT# 3051302; BLUNT TIP SCREW, 4X46MM, ITEM# 47248604640, LOT# 3073796; BLUNT TIP SCREW, 4X48MM, ITEM# 47248604840, LOT# 3054547; TORQUE LIMITING HANDLE, ITEM# 27923, LOT# N/A. REPORT SOURCE- FOREIGN: JAPAN. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0009613350 -2022 -00596; 0009613350 -2022 -00599; 0009613350 -2022 -00607. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS; HOWEVER, AN INVESTIGATION OF THE REPORTED EVENT IS IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL MEDWATCH 3500A WILL BE SUBMITTED.
(B)(4). THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL AND/OR CORRECTED INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED. AS NO PRODUCTS WERE RETURNED, VISUAL AND DIMENSIONAL EVALUATIONS COULD NOT BE PERFORMED. REVIEW OF THE DEVICE HISTORY RECORDS IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING. DEVICES ARE USED FOR TREATMENT. AN UNDATED X-RAY IMAGE OF THE AFFECTED HUMERUS IS AVAILABLE. THE IMAGE SHOWS WHAT APPEARS TO BE A BACKING-OUT OF THE MOST PROXIMAL SCREW. HOWEVER, WITHOUT ANY FURTHER X-RAYS FOR COMPARISON PROVIDED, NO FURTHER ASSESSMENT CAN BE MADE. NO FURTHER MEDICAL DOCUMENTS SUCH AS SURGICAL REPORTS WERE PROVIDED. BASED ON THE INVESTIGATION IT COULD BE ASSUMED THAT FURTHER POSSIBLE CONTRIBUTING FACTORS TO THE MIGRATION OF THE SCREW MIGHT BE MULTIFACTORIAL RELATED TO EITHER PATIENT CONDITION, BEHAVIOUR OR IMPLANTATION PROCEDURE. IF AND TO WHAT EXTENT ANY OF THESE ASPECTS MAY HAVE INFLUENCED THE MIGRATION OF THE SCREW REMAINS UNKNOWN. AN ADDITIONAL DEEPER INVESTIGATION WAS PERFORMED WHICH IDENTIFIED THE DESIGN LIMITATION OF THE CORELOCK MECHANISM OF THE AFFIXUS NAIL AS A POTENTIAL CONTRIBUTING FACTOR. AS PART OF THE DEEPER INVESTIGATION, ASCIENTIFIC LITERATURE SEARCH WAS CONDUCTED AND A SYSTEMATIC REVIEW OF THE OUTCOME OF INTRAMEDULLARY NAILING FOR ACUTE PROXIMAL HUMERUS FRACTURE SHOWED THAT SCREW MIGRATION AND PERFORATION INTO THE JOINT IS THE SECOND MOST COMMON COMPLICATION FOLLOWING SECONDARY LOSS OF REDUCTION. IN ADDITION, SCIENTIFIC LITERATURE OF COMPETITOR AND PREDICATE DEVICES WERE ASSESSED. THIS REVIEW HAS SHOWN THAT THE AFFIXUS. NATURAL NAIL PROXIMAL HUMERAL SYSTEM PERFORMS BETTER AND/OR IN EQUAL RANGE IN COMPARISON WITH LEGALLY MARKETED SIMILAR DEVICES. THEREFORE, NO DESIGN CHANGES WERE CONDUCTED. IN CONCLUSION, AS THE CAUSE MAY BE MULTIFACTORIAL AN EXACT ROOT CAUSE COULD NOT BE IDENTIFIED FOR THE MIGRATION OF THE SCREW. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS REPORTED THAT 2-MONTH POST-IMPLANTATION, A SCREW BACKED OUT FROM ITS CORRECT POSITION. NO REVISION SURGERY IS CURRENTLY SCHEDULED. DUE DILIGENCE IS IN PROGRESS FOR THIS COMPLAINT; TO DATE NO ADDITIONAL INFORMATION OR PRODUCT HAS BEEN RECEIVED.
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 322128 | BLUNT TIP SCREW, 4X48MM | ROD, FIXATION | HSB | ZIMMER MANUFACTURING SWITZERLAND GMBH | N/A | 3073798 | 00889024505483 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Prefer Not To Disclose | Other |