ANEURX STENT GRAFT WITH XCELERANT DELIVERY SYSTEM
Report
- Report Number
- 2953200-2010-00046
- Event Type
- Death
- Date Received
- January 13, 2010
- Date of Event
- December 11, 2006
- Report Date
- February 21, 2008
- Manufacturer
- MEDTRONIC CARDIOVASCULAR
- Product Code
- MIH
- PMA / PMN Number
- P990020
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- PHYSICIAN
Narratives
(INTERVENTION REQUIRED) - EVALUATION RESULTS: DEATH; CAUSE OF DEATH NOT PROVIDED.
AN ANEURX STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM APPROX 6 YEARS AGO. THE PT WAS ENROLLED IN A STUDY. ANEURYSM AND VESSEL MORPHOLOGY AT THE TIME OF IMPLANT WERE NOT REPORTED. THE PT'S MEDICAL HISTORY IS NOTABLE FOR MYOCARDIAL INFARCTION, HYPERTENSION, DIABETES, RENAL INSUFFICIENCY, AND CHRONIC OBSTRUCTION PULMONARY DISEASE. IT WAS REPORTED THAT A FOLLOW-UP CT SCAN APPROX 23 MONTHS POST-IMPLANT REVEALED A 1 CM DISTAL MIGRATION OF THE ANEURX BIFURCATED STENT GRAFT WITH A LARGE TYPE I ENDOLEAK FILLING THE ANEURYSM SAC (MFR# 2953200-2010-00045). THE PHYSICIAN ELECTED TO INTERVENE BY RELINING THE RIGHT ILIAC LIMB WITH AN ANEURX ILIAC LIMB AND ALSO IMPLANTING AN INVESTIGATIONAL TALENT AORTIC CUFF. THE PT'S POST-OPERATIVE COURSE WAS UNREMARKABLE, AND THE 1 MONTH FOLLOW UP CT SHOWED THAT THE GRAFTS WERE INTACT WITHOUT EVIDENCE OF KINKING OF ENDOLEAKS. APPROX 1 YEAR POST-INTERVENTION, IT WAS REPORTED THAT THE ANEURX DEVICE HAD MIGRATED DISTALLY DUE TO ANTERIOR ANGULATION WITH POOR PROXIMAL APPOSITION. IN ADDITION, THE ANEURYSM SAC HAD ENLARGED FROM 5.5 CM TO 6.1 CM, DUE TO A TYPE III ENDOLEAK, SECONDARY TO MINIMAL OVERLAP BETWEEN THE ANEURX BIFURCATED STENT GRAFT AND THE TALENT CUFF. THE PHYSICIAN ELECTED TO INTERVENE BY IMPLANTING 3 AORTIC CUFF STENT GRAFTS FROM ANOTHER MFR, WITH SUCCESSFUL RESULTS, FROM THE RENAL ARTERIES TO THE BIFURCATION OF THE ANEURX BIFURCATED STENT GRAFT. THE ANEURX LIMB WAS ALSO SUCCESSFULLY RELINED WITH AN ILIAC LIMB FROM ANOTHER MFR. THE INVESTIGATOR ASSESSED THAT THE MIGRATION WAS RELATED TO THE ANEURX BIFURCATED GRAFT BUT NOT RELATED TO THE PROCEDURE. APPROXIMATELY 34 MONTHS POST-INTERVENTION, THE PT EXPIRED. THE CAUSE OF DEATH HAS NOT BEEN PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANEURX STENT GRAFT WITH XCELERANT DELIVERY SYSTEM | MIH | MEDTRONIC CARDIOVASCULAR | NA | 670412 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Death |