LIGACLIP ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER
Report
- Report Number
- 1527736-1998-00798
- Event Type
- Injury
- Date Received
- March 20, 1998
- Date of Event
- January 14, 1998
- Report Date
- February 19, 1998
- Manufacturer
- ETHICON ENDO-SURGERY, INC.
- Product Code
- GDO
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
TRACKING #.49799. EES #.981269/A. THE PT UNDERWENT A LAPAROSCOPIC CHOLECYSTECTOMY IN 1/1998 AND ONE WEEK POST OPERATIVELY DEVELOPED BILE PERITONITIS. THEY WERE UNABLE TO PLACE A STENT IN THE LOCAL HOSP AND THE PT WAS SENT TO ANOTHER INSTITUTION WHERE THE STENT WAS PLACED. "ERCP" REVEALED A LEAK FROM THE END OF THE DUCT. A CLIP WAS PRESENT BUT MAY NOT HAVE BEEN COMPLETELY CLOSED. LATER THE PT REQUIRED EXPLORATORY LAPROTOMY WITH DRAINAGE OF A LESSER SAC ABSCESS. CO DISCUSSED PROPER TECHNIQUE IN APPLYING CLIPS. THE PHYSICIAN STATES THE HOSP TESTS THE CLIP APPLIER BY FIRING THE CLIP ON A TOWEL PRIOR TO ITS USE. FACILITY EXPLAINED THAT THIS WAS SOMETHING CO HAD NOT VALIDATED AND COULD NOT RECOMMEND. SPECIFICALLY, FACILITY INDICATED THAT FIRING THE INSTRUMENT IN AN INAPPROPRIATE FASHION COULD LEAD TO DAMAGED JAWS AND NON FUNCTIONAL CLIPS.
IT WAS REPORTED AFTER A LAPAROSCOPIC CHOLECYSTECTOMY THE PT HAD POST-OP DISCOMFORT. AN ERCP FAILED TWO TIMES DUE TO "ANATOMICAL LIMITATIONS." THE PT WAS SENT TO THE MEDICAL CENTER FOR AN EVALUATION AND STENT. THE DRS AT MEDICAL CENTER SAID THE CYSTIC DUCT WAS LEAKING BILE DUE TO FAILURE OF THE CLIPS BETWEEN THE CYSTIC DUCT AND COMMON DUCT. THE SURGEON AND TECH MENTIONED THAT THE CLIPS WERE THE ANODIZED BLUE COLOR NOT THE NORMAL SILVER TITANIUM FINISH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGACLIP ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER | ENDOSCOPIC CLIP APPLIER | GDO | ETHICON ENDO-SURGERY, INC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other| R |