FDA Adverse Event Malfunction Summary report: N

BD PHASEAL¿ SPIKE CONNECTOR

MDR report key: 15811316 · Received November 16, 2022

Report

Report Number
3003152976-2022-00519
Event Type
Malfunction
Date Received
November 16, 2022
Date of Event
October 19, 2022
Report Date
November 28, 2022
Manufacturer
BECTON DICKINSON, S.A.
Product Code
ONB
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL YES, D10: RETURNED TO MANUFACTURER ON: 28-NOV-2022. H6: INVESTIGATION SUMMARY ONE SAMPLE AND MULTIPLE PHOTOS WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, IT IS OBSERVED THAT THE SPIKE IS BROKEN, A PIECE OF THE SPIKE THREAD REMAINED STUCK ON THE CONNECTOR THREAD. TRACES OF THE SOLVENT USED TO BOND THE SPIKE TO THE CONNECTOR WERE OBSERVED. PRODUCT UNDERGOES INSPECTIONS THROUGHOUT THE MANUFACTURING PROCESS TO ENSURE THE QUALITY AND FUNCTIONALITY OF THE DEVICE, INCLUDING FORCE TESTING TO ENSURE PROPER ASSEMBLY OF THE DEVICE. A DEVICE HISTORY REVIEW WAS PERFORMED FOR REPORTED LOT 2201210, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. BASED ON THE SAMPLE EVALUATION IT WAS DETERMINED THE BREAKAGE LIKELY OCCURRED DURING USE, WHEN INSERTING THE DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD PHASEAL¿ SPIKE CONNECTOR EXPERIENCED A BROKEN SPIKE TIP. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: UPON ATTEMPT TO ENGAGE N35J WITH A SYRINGE, THE BONDING OF THE SPIKE AND C35J ON C180J WAS EASILY BROKEN.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD PHASEAL¿ SPIKE CONNECTOR EXPERIENCED A BROKEN SPIKE TIP. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: UPON ATTEMPT TO ENGAGE N35J WITH A SYRINGE, THE BONDING OF THE SPIKE AND C35J ON C180J WAS EASILY BROKEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
375265 BD PHASEAL¿ SPIKE CONNECTOR INTRAVASCULAR ADMINISTRATION SET ONB BECTON DICKINSON, S.A. 2201210

Patients

Seq Age Sex Outcome Treatment
1 Unknown