BD PHASEAL¿ L CONNECTOR C90J
Report
- Report Number
- 3003152976-2022-00518
- Event Type
- Malfunction
- Date Received
- November 16, 2022
- Date of Event
- October 24, 2022
- Report Date
- December 15, 2022
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- ONB
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL YES. D10: RETURNED TO MANUFACTURER ON: 15-DEC-2022. H6: INVESTIGATION SUMMARY: ONE SAMPLE WAS PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. AFTER VISUAL INSPECTION, IT WAS OBSERVED THAT THE MEMBRANE WAS MISSING AND NEVER ASSEMBLED ONTO THE CONNECTOR BODY, THEREFORE THE SYSTEM WAS OPEN AND LEAKED WHEN USED. A DEVICE HISTORY REVIEW WAS PERFORMED FOR LOT 2202204, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. ASSEMBLY OF THE MEMBRANE IS PERFORMED MANUALLY. A DETECTION SYSTEM IS USED TO VERIFY THE PROPER WELDING AND LOCATION OF THE MEMBRANE. ALL QUALITY RECORDS VERIFY THE INSPECTIONS WERE PERFORMED ACCORDING TO PROCEDURE. WHILE WE CANNOT IDENTIFY A DIRECT ISSUE, IT WAS DETERMINED THIS INCIDENT IS DUE TO THE OPERATOR NOT WELDING THE MEMBRANE ONTO THE CONNECTOR BODY AND THE DETECTION SYSTEM ALSO NOT IDENTIFYING/REJECTING THE IMPACTED PIECE.
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE BD PHASEAL¿ L CONNECTOR C90J EXPERIENCED LEAKAGE BETWEEN THE INJECTOR AND CONNECTOR. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DURING ADMINISTRATION OF OXALIPLATIN, THE MEMBRANE ON L-CONNECTOR WAS DISCONNECTED, RESULTING IN LEAKAGE.
IT WAS REPORTED THAT THE BD PHASEAL¿ L CONNECTOR C90J EXPERIENCED LEAKAGE BETWEEN THE INJECTOR AND CONNECTOR. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DURING ADMINISTRATION OF OXALIPLATIN, THE MEMBRANE ON L-CONNECTOR WAS DISCONNECTED, RESULTING IN LEAKAGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 374457 | BD PHASEAL¿ L CONNECTOR C90J | INTRAVASCULAR ADMINISTRATION SET | ONB | BECTON DICKINSON, S.A. | 2202204 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |