FDA Adverse Event Other Summary report: N

EVICEL (FACTOR I - FIBRINOGEN)

MDR report key: 1581097 · Received January 16, 2010

Report

Report Number
3003183625-2010-00001
Event Type
Other
Date Received
January 16, 2010
Date of Event
December 17, 2009
Report Date
January 15, 2010
Manufacturer
OMRIX BIOPHARMACEUTICALS LTD., SHEBA HOSPITAL
Product Code
BXX
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

A SERIOUS SPONTANEOUS REPORT REGARDING A (B)(6) MALE PATIENT WHO DEVELOPED AIR EMBOLISM AFTER BEING TREATED WITH EVICEL VIA AN AIR PRESSURE DEVICE, DURING SURGERY FOR A LAPAROSCOPIC PARTIAL NEPHRECTOMY. THE CASE WAS REPORTED BY A PHYSICIAN VIA A SALES REP. THE PT WAS REPORTED TO BE RELATIVELY FIT AND WELL BUT HAD A HISTORY OF WOLFF PARKINSON WHITE SYNDROME. HE WAS CURRENTLY BEING TREATED MEDICALLY; HOWEVER, THE MEDICATION FOR THIS SYNDROME AND OTHER CONCOMITANT MEDICATIONS WAS NOT PROVIDED. IT WAS REPORTED THAT THE PT HAD BEEN VERY STABLE THROUGHOUT THE SURGERY. THE RENAL ARTERY HAD BEEN CLAMPED PRIOR TO EVICEL APPLICATION. WITHIN 2-3 MINS OF APPLICATION OF EVICEL USING THE GAS INJECTOR (WITH THE TIP WITHIN 1 CM TO THE SURFACE OF THE RESECTED RENAL MARGIN), THE PT BECAME VERY UNSTABLE. THERE WAS A SUDDEN DROP IN THE PT'S OXYGEN SATURATION FROM 99% TO 85%. THIS WAS ALSO ASSOCIATED WITH A FALL BY 30-40% IN THE PT'S END-TIDAL CO2 (ETCO2). HIS BLOOD PRESSURE REMAINED UNCHANGED AND HIS HEART RATE WAS INITIALLY UNCHANGED, THEN FELL AND HIS RHYTHM CHANGED FROM SINUS TO BIGEMINY. THE TREATING SURGEON FELT THIS COULD BE EITHER A EQUIPMENT PROBLEM, LUNG COLLAPSE, TUBE MISPLACEMENT / DISLODGEMENT, ANAPHYLAXIS, FIBRIN EMBOLUS, AIR EMBOLUS OR A CARDIAC CONDITION. THE PT WAS GIVEN 100% OXYGEN AND WAS THEN OFF THE VENTILATOR AND MANUALLY VENTILATED. LUNG COMPLIANCE WAS NORMAL. AUSCULTATION OF THE CHEST WAS NORMAL WITH GOOD AIR ENTRY THROUGHOUT AND NO WHEEZE OR SIGNS OF PNEUMOTHORAX. THE ANAESTHETIST RESPONSIBLE FOR THE PT, NOTICED THE EMBOLI AND STARTED TO DEFLATE THE INSUFFLATED GAS FROM THE SURGICAL FIELD. AN EXTERNAL JUGULAR VEIN LINE WAS ALREADY IN SITU AND THIS WAS ASPIRATED, INITIALLY BLOOD THEN AIR WAS ASPIRATED. AT THIS TIME, THE SURGEON WAS NOT SURE WHETHER THIS WAS ENTRAINED AIR FROM THE SYRINGE BECOMING LOOSE BUT IN RETROSPECT, THE SURGEON FELT THIS WAS MOST LIKELY TO BE TRUE ASPIRATION. THIS AIR AMOUNTED TO APPROX 15 MLS. AT FIRST, THE SURGEON DID NOT HAVE TIME TO LISTEN FOR A MILLWHEEL MURMUR. INTRAVENOUS GLYCOPYRROLATE AT 200MCG WAS GIVEN SLOWLY. THIS HAD GOOD EFFECT FOR THE BIGEMINY. THE PT RETURNED TO SINUS RHYTHM. OVER THE NEXT 3-5 MINUTES, THERE WAS A STEADY RISE IN THE PT'S OXYGEN SATURATION AND ETCO2. BLOOD SAMPLES WERE TAKEN FOR MAST CELL TRYPTASE TESTS TO EXCLUDE ON ALLERGIC CAUSE BUT THE HISTORY STRONGLY SUGGESTED GAS EMBOLUS. RESULTS OF THE TRYPTASE TEST WERE NOT PROVIDED. IT WAS REPORTED THAT THE PT DID MAKE A FULL AND UNEVENTFUL RECOVERY WITH NO SEQUELAE. THE REPORTER CONSIDERED THE EVENT TO BE RELATED TO THE PRESSURE OF APPLICATION. THE COMPANY PHYSICIAN ASSESSED THE CAUSALITY AS RELATED TO THE USE OF THE PRESSURE REGULATOR. DETAILS OF THE DEVICE ARE : 5 ML DEVICE: (B)(4); 1/2ML: (B)(4); 35 CM: (B)(4); 45CM: (B)(4); PRESSURE REGULATOR: (B)(4). APPLICATION DEVICE: (B)(4); PRESSURE REGULATOR: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EVICEL (FACTOR I - FIBRINOGEN) PRESSURE REGULATOR DEVICE BXX OMRIX BIOPHARMACEUTICALS LTD., SHEBA HOSPITAL PLS SEE NARRATIVE PLS SEE NARRATIVE

Patients

Seq Age Sex Outcome Treatment
1 51 YR Other