FDA Adverse Event Malfunction Summary report: N

BD¿ LUER SLIP TIP SYRINGE WITH ATTACHED NEEDLE

MDR report key: 15801828 · Received November 15, 2022

Report

Report Number
1213809-2022-01002
Event Type
Malfunction
Date Received
November 15, 2022
Date of Event
October 24, 2022
Report Date
January 17, 2023
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096238
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL YES, RETURNED TO MANUFACTURER ON: 19-DEC-2022. H6: INVESTIGATION SUMMARY TWELVE SAMPLES WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THE REPORTED ISSUE WAS NOT CONFIRMED UPON INSPECTION OF THE SAMPLES. NONE OF THE SAMPLES SHOWED ANY SIGNS OF THE REPORTED DEFECT. SINCE THE REPORTED DEFECT WAS NOT CONFIRMED A MANUFACTURING ROOT CAUSE COULD NOT BE DETERMINED. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 1113470. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.

Additional Manufacturer Narrative · 0

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD¿ LUER SLIP TIP SYRINGE WITH ATTACHED NEEDLE EXPERIENCED BEING UNABLE TO REMOVE THE NEEDLE SHIELD, AND NEEDLE AND SYRINGE SEPARATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: SPOUSE REPORTED SHE HAS TO USE "PLYERS" TO REMOVE NEEDLE SHIELD. STATED, SHE HAS TO PULL REALLY HARD AND WHEN THE NEEDLE SHIELD FINALLY COMES OFF, THE NEEDLE COMES OFF WITH THE SHIELD.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD¿ LUER SLIP TIP SYRINGE WITH ATTACHED NEEDLE EXPERIENCED BEING UNABLE TO REMOVE THE NEEDLE SHIELD, AND NEEDLE AND SYRINGE SEPARATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: SPOUSE REPORTED SHE HAS TO USE "PLIERS" TO REMOVE NEEDLE SHIELD. STATED, SHE HAS TO PULL REALLY HARD AND WHEN THE NEEDLE SHIELD FINALLY COMES OFF, THE NEEDLE COMES OFF WITH THE SHIELD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2829616 BD¿ LUER SLIP TIP SYRINGE WITH ATTACHED NEEDLE PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 309623 1113470 30382903096238

Patients

Seq Age Sex Outcome Treatment
1 Unknown