SYNCHRON® LX®I 725 CLINICAL SYSTEM
Report
- Report Number
- 2050012-2010-00007
- Event Type
- Malfunction
- Date Received
- January 19, 2010
- Date of Event
- December 14, 2009
- Report Date
- January 19, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- CHH
- PMA / PMN Number
- K023049
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
QC WAS WITHIN CUSTOMER SPECIFICATIONS BEFORE THE EVENT. A FIELD SERVICE ENGINEER (FSE) WAS ON-SITE AND REPAIRED, REPLACED OR ALIGNED THE SAMPLE AND REAGENT SYRINGE SYSTEM COMPONENTS, THE SAMPLE AND REAGENT PROBE SYSTEM COMPONENTS, AND THE SAMPLE AND REAGENT MIXER SYSTEM COMPONENTS. DUE TO A POTENTIAL FOR THIS EVENT TO RECUR UNKNOWINGLY, THERE IS POTENTIAL THAT A PIVOTAL ASSAY COULD BE ERRONEOUSLY REPORTED AND TREATMENT DECISION MAY BE AFFECTED. A MALFUNCTION WILL BE ASSUMED FOR THE PURPOSE OF THIS REPORT.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING MULTIPLE FALSELY HIGH CHOLESTEROL, TRIGLYCERIDE AND HDL-CHOLESTEROL RESULTS GENERATED BY THE SYNCHRON LXI 725 CLINICAL SYSTEM. THE INITIAL CHOLESTEROL, TRIGLYCERIDE AND HDL RESULTS WERE IN THE RANGE 230-452, 108-1046 AND 39-111 UNITS RESPECTIVELY; AND THE REPEAT RESULTS WERE IN THE RANGE 150-299, 66-615 AND 30-72 UNITS RESPECTIVELY. THE ERRONEOUS RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE LABORATORY BELIEVES THAT ONE OR TWO PATIENTS WERE PRESCRIBED MEDICATION BASED ON THE ERRONEOUS RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON® LX®I 725 CLINICAL SYSTEM | CLINICAL CHEMISTRY ANALYZER | CHH | BECKMAN COULTER INC. | LX®I 725 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |