FDA Adverse Event Injury Summary report: N

RENESSA RF SYSTEM

MDR report key: 1580072 · Received January 12, 2010

Report

Report Number
3003647794-2010-00083
Event Type
Injury
Date Received
January 12, 2010
Date of Event
October 1, 2009
Report Date
January 1, 2010
Manufacturer
NOVASYS MEDICAL
Product Code
NVJ
PMA / PMN Number
K042132
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
ND, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

NO DEVICE MALFUNCTIONED OCCURRED. WORSENING INCONTINENCE IS A KNOWN RISK FOLLOWING ANY ANTI-INCONTINENCE PROCEDURE. FOLLOWING RENESSA, SOME PATIENTS HAVE EXPERIENCED A WORSENING OF INCONTINENCE 3-4 WEEKS FOLLOWING THE TREATMENT AT THE TIME THAT THE URETHRAL SUBMUCOSAL TISSUE IS UNDERGOING COLLAGEN REMODELING FOLLOWING URETHRAL SUBMUCOSAL COLLAGEN DENATURATION DURING THE TREATMENT. SYMPTOMS CAN INCLUDE URINARY TRACT INFECTION, OVERFLOW URINARY INCONTINENCE, OVERACTIVE BLADDER, OR INTRINSIC SPHINCTER DEFICIENCY. ALL OF THESE CONDITIONS ARE TREATABLE. THE MANUFACTURER HAS EVALUATED DEVICES ON AN ONGOING BASIS TO DETERMINE IF ANY DEVICE RELATED MALFUNCTION MAY HAVE CONTRIBUTED TO THE OCCURRENCE OF THIS KNOWN ADVERSE EVENT. TO DATE, NO DEVICE MALFUNCTION HAS BEEN FOUND TO CONTRIBUTE TO THE EVENT, BUT INVESTIGATIONS CONTINUE. THE MANUFACTURER HAS CREATED PROCEDURAL AIDS TO EMPHASIZE APPROPRIATE PT SELECTION, HAS SIMPLIFIED THE STEPS OF THE PROCEDURE, MADE CHANGES TO THE INSTRUCTIONS FOR USE AND HAD MADE MINOR DEVICE CHANGES TO FACILITATE EASE OF USE BY THE PHYSICIAN. THE MANUFACTURER CONTINUES TO EVALUATE THE EFFECTIVENESS OF THESE STEPS TO MINIMIZE THE OCCURRENCE OF THIS SIDE EFFECT.

Description of Event or Problem · 1

PT BEGAN TO HAVE WORSENING INCONTINENCE REQUIRING CONSTANT CHANGING OF PADS 4-5 WEEKS AFTER THE PROCEDURE. PT TO HAVE A MIDURETHAL SLING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RENESSA RF SYSTEM NVJ NOVASYS MEDICAL PR0918

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention RENESSA RF GENERATOR