COOL POINT¿ IRRIGATION PUMP PRESSURE ALARM ADAPTER CABLE
Report
- Report Number
- 2030404-2022-00070
- Event Type
- Malfunction
- Date Received
- November 14, 2022
- Date of Event
- October 26, 2022
- Report Date
- January 13, 2023
- Manufacturer
- ST. JUDE MEDICAL
- Product Code
- OAD
- PMA / PMN Number
- P060019
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
ADDITIONAL INFORMATION: G3, H2, H3 THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE SINCE THE DEVICE WAS NOT RETURNED FOR ANALYSIS. REVIEW OF THE DEVICE HISTORY RECORD WAS NOT POSSIBLE AS THE LOT NUMBER IS UNKNOWN. BASED ON THE INFORMATION RECEIVED, THE CAUSE OF THE REPORTED ERROR MESSAGE AND SUBSEQUENT DELAY REMAINS UNKNOWN.
RELATED MANUFACTURING REF: 2030404-2022-00068 AND 2030404-2022-00069. DURING THE ATRIAL TACHYCARDIA ABLATION, THERE WAS A PROCEDURE DELAY DUE TO PRESSURE AND BUBBLE ERRORS DUE TO THE PRESSURE ADAPTOR CABLE AND TWO TUBING SETS. WHILE PREPPING THE COOL POINT PUMP TUBING FOR THE TACTICATH SE THERE WAS A REPEATING PRESSURE ERROR ONCE THE CATHETER WAS HOOKED UP. THE TUBING HAD ALREADY BEEN FLUSHED ON 60 ML/MIN AND HAD RUN WITHOUT ISSUE. THE CATHETER WAS THEN DISCONNECTED AND IT WAS ATTEMPTED TO FLUSH IT, BUT THE PRESSURE ERROR REOCCURRED. THE COOL POINT PUMP WAS TURNED OFF AND ON AND RECONNECTED TO THE TUBING MULTIPLE TIMES. ONCE THE PRESSURE ADAPTOR CABLE WAS REPLACED THE PRESSURE ERROR WENT AWAY. THE PUMP RAN AT 60, BUT INSTEAD WAS GIVING THE BUBD ERROR WHEN ATTEMPTING TO RUN AT 2ML/MIN. NO BUBBLES WERE OBSERVED SO A NEW TUBING SET WAS EXCHANGED, BUT THE BUBD ERROR DID NOT RESOLVE. A LEAK WAS NOTICED WITHIN THE TWIST CONNECTION OF THE TUBING AND DESPITE TIGHTENING THE LEAK DID NOT STOP. A THIRD TUBING SET WAS USED AND THERE WERE NO FURTHER ISSUES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO ADVERSE CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1199005 | COOL POINT¿ IRRIGATION PUMP PRESSURE ALARM ADAPTER CABLE | CARDIAC ABLATION PERCUTANEOUS CATHETER | OAD | ST. JUDE MEDICAL | H701666 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Female | COOL POINT¿ IRRIGATION PUMP TUBING SET| COOL POINT¿ IRRIGATION PUMP TUBING SET |