FDA Adverse Event Malfunction Summary report: N

COOL POINT¿ IRRIGATION PUMP PRESSURE ALARM ADAPTER CABLE

MDR report key: 15794887 · Received November 14, 2022

Report

Report Number
2030404-2022-00070
Event Type
Malfunction
Date Received
November 14, 2022
Date of Event
October 26, 2022
Report Date
January 13, 2023
Manufacturer
ST. JUDE MEDICAL
Product Code
OAD
PMA / PMN Number
P060019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: G3, H2, H3 THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE SINCE THE DEVICE WAS NOT RETURNED FOR ANALYSIS. REVIEW OF THE DEVICE HISTORY RECORD WAS NOT POSSIBLE AS THE LOT NUMBER IS UNKNOWN. BASED ON THE INFORMATION RECEIVED, THE CAUSE OF THE REPORTED ERROR MESSAGE AND SUBSEQUENT DELAY REMAINS UNKNOWN.

Description of Event or Problem · 0

RELATED MANUFACTURING REF: 2030404-2022-00068 AND 2030404-2022-00069. DURING THE ATRIAL TACHYCARDIA ABLATION, THERE WAS A PROCEDURE DELAY DUE TO PRESSURE AND BUBBLE ERRORS DUE TO THE PRESSURE ADAPTOR CABLE AND TWO TUBING SETS. WHILE PREPPING THE COOL POINT PUMP TUBING FOR THE TACTICATH SE THERE WAS A REPEATING PRESSURE ERROR ONCE THE CATHETER WAS HOOKED UP. THE TUBING HAD ALREADY BEEN FLUSHED ON 60 ML/MIN AND HAD RUN WITHOUT ISSUE. THE CATHETER WAS THEN DISCONNECTED AND IT WAS ATTEMPTED TO FLUSH IT, BUT THE PRESSURE ERROR REOCCURRED. THE COOL POINT PUMP WAS TURNED OFF AND ON AND RECONNECTED TO THE TUBING MULTIPLE TIMES. ONCE THE PRESSURE ADAPTOR CABLE WAS REPLACED THE PRESSURE ERROR WENT AWAY. THE PUMP RAN AT 60, BUT INSTEAD WAS GIVING THE BUBD ERROR WHEN ATTEMPTING TO RUN AT 2ML/MIN. NO BUBBLES WERE OBSERVED SO A NEW TUBING SET WAS EXCHANGED, BUT THE BUBD ERROR DID NOT RESOLVE. A LEAK WAS NOTICED WITHIN THE TWIST CONNECTION OF THE TUBING AND DESPITE TIGHTENING THE LEAK DID NOT STOP. A THIRD TUBING SET WAS USED AND THERE WERE NO FURTHER ISSUES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO ADVERSE CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1199005 COOL POINT¿ IRRIGATION PUMP PRESSURE ALARM ADAPTER CABLE CARDIAC ABLATION PERCUTANEOUS CATHETER OAD ST. JUDE MEDICAL H701666

Patients

Seq Age Sex Outcome Treatment
1 Female COOL POINT¿ IRRIGATION PUMP TUBING SET| COOL POINT¿ IRRIGATION PUMP TUBING SET