FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V MTX 3.7MM 3.5MM 11.5MM

MDR report key: 15793184 · Received November 14, 2022

Report

Report Number
0002023141-2022-02872
Event Type
Injury
Date Received
November 14, 2022
Date of Event
October 19, 2022
Report Date
March 13, 2023
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019515
PMA / PMN Number
K061410
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT NUMBER - (B)(4). ADDITIONAL PMA/510(K) NUMBER ¿ K013227. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR BONE LOSS AND INFECTION EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR BONE LOSS AND INFECTION HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF MANUFACTURING OR DESIGN DEFECTS THAT MIGHT LEAD TO BONE LOSS AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NON-EXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: H6: TYPE OF INVESTIGATION CODE WAS ADDED: 3331. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1233260. IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. STERILIZATION RECORD OP# 150 - STR2 WAS REVIEWED AND VERIFIED TO HAVE PASSED ALL STERILIZATION ACTIVITIES WITH NO ISSUES OR NONCONFORMITIES IDENTIFIED. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1233260 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. AS DOCUMENTED IN THE SUMMARY INVESTIGATIONS, CONTRIBUTING FACTORS FOR THE REPORTED EVENT LIKELY EXIST OUTSIDE OF ZIMMER BIOMET CONTROL, INCLUDING THOSE RELATED TO PATIENT BIOLOGICAL FACTORS/CONDITION AND SURGICAL TECHNIQUE. BASED ON THE SUMMARY INVESTIGATIONS, NO MALFUNCTION OCCURRED UPON INVESTIGATION. THE REPORTED EVENTS REMAIN NON-VERIFIABLE AS THEY ARE A MEDICAL CONDITION.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE IMPLANT AT TOOTH LOCATION #45 WAS REMOVED DUE TO INFECTION AND BONE LOSS. SYMPTOMS AS A RESULT OF THE EVENT: PAIN.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2705113 IMPL TAPERED SCR-V MTX 3.7MM 3.5MM 11.5MM DENTAL IMPLANT DZE ZIMMER DENTAL TSVB11 1233260 00889024019515

Patients

Seq Age Sex Outcome Treatment
1 39 YR Female Required Intervention