FDA Adverse Event Malfunction Summary report: N

ARCHITECT ANTI-HBS REAGENT KIT

MDR report key: 15788415 · Received November 14, 2022

Report

Report Number
3008344661-2022-00156
Event Type
Malfunction
Date Received
November 14, 2022
Date of Event
November 5, 2022
Report Date
January 23, 2023
Manufacturer
ABBOTT IRELAND
Product Code
LOM
PMA / PMN Number
P050051
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 7C18 HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 1L82.

Additional Manufacturer Narrative · 0

AFTER FURTHER EVALUATION, THE SUSPECT MEDICAL DEVICE WAS CHANGED FROM ARCHITECT ANTI-HBS, 07C18-78, ABBOTT IRELAND TO ARCHITECT I2000SR PROCESSING MODULE, 03M74-02, ABBOTT LABORATORIES IRVING. MDR NUMBER 3016438761-2023-00034-00 HAS BEEN SUBMITTED AND ALL FURTHER INFORMATION WILL BE DOCUMENTED UNDER THAT MDR NUMBER. H3 OTHER TEXT : AFTER FURTHER EVALUATION, THE SUSPECT MEDICAL DEVICE WAS CHANGED FROM ARCHITECT ANTI-HBS, 07C18-78, ABBOTT IRELAND TO ARCHITECT I2000SR PROCESSING MODULE, 03M74-02, ABBOTT LABORATORIES IRVING. MDR NUMBER 3016438761-2023-00034-00 HAS BEEN SUBMITTED AND ALL FURTHER INFORMATION WILL BE DOCUMENTED UNDER THAT MDR NUMBER.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED A FALSELY ELEVATED ARCHITECT ANTI-HBS RESULT FOR ONE SAMPLE. THE FOLLOWING DATA WAS PROVIDED: INITIAL RESULT = 74.11 MIU/ML (PROTECTIVE), REPEAT = 1.76 MIU/ML (NOT PROTECTIVE). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1830200 ARCHITECT ANTI-HBS REAGENT KIT TEST, HEPATITIS B (B CORE, BE ANTIGEN, BE ANTIBODY, B CORE IGM) LOM ABBOTT IRELAND 34344FN01

Patients

Seq Age Sex Outcome Treatment
1 Unknown ARC I2000SR INST, 03M74-02, (B)(6)| ARC I2000SR INST, 03M74-02, (B)(6)