FDA Adverse Event Malfunction Summary report: N

FS OPT H KET-CODING

MDR report key: 15786754 · Received November 14, 2022

Report

Report Number
2954323-2022-41569
Event Type
Malfunction
Date Received
November 14, 2022
Date of Event
October 24, 2022
Report Date
November 23, 2022
Manufacturer
ABBOTT DIABETES CARE INC
Product Code
NBW
PMA / PMN Number
K040814
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE PRODUCT HAS BEEN REQUESTED BACK FOR INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 0

REPEATED ATTEMPTS BY ADC TO RETRIEVE THE PRODUCT WERE UNSUCCESSFUL AND/OR THE CUSTOMER DISCARDED THE PRODUCT. NO PRODUCT HAS BEEN RETURNED. EXTENDED INVESTIGATION HAS BEEN PERFORMED FOR THE REPORTED COMPLAINT AND THERE WAS NO INDICATION THAT THE PRODUCT DID NOT MEET SPECIFICATION. DHRS FOR THE XCEED OPTIUM METER WAS REVIEWED AND THE DHRS SHOWED THE XCEED OPTIUM METER PASSED ALL TESTS PRIOR TO RELEASE. DHRS FOR THE PRECISION STRIPS WAS REVIEWED AND THE DHRS SHOWED THE PRECISION STRIPS PASSED ALL TESTS PRIOR TO RELEASE. RETAIN TESTING WAS PERFORMED FOR PRECISION STRIPS, AND ALL UNITS PERFORMED IN SPECIFICATION AND PASSED. IF THE PRODUCT IS RETURNED, A PHYSICAL INVESTIGATION WILL BE PERFORMED AND A FOLLOW-UP REPORT SUBMITTED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Description of Event or Problem · 0

A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Description of Event or Problem · 0

A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
353148 FS OPT H KET-CODING BLOOD GLUCOSE MONITORING SYSTEM NBW ABBOTT DIABETES CARE INC 98814-65 4500185979

Patients

Seq Age Sex Outcome Treatment
1 Unknown