INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2022-20047
- Event Type
- Injury
- Date Received
- November 11, 2022
- Date of Event
- October 13, 2022
- Report Date
- December 24, 2022
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FTR
- UDI-DI
- 10888628007130
- PMA / PMN Number
- P020056
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE AND BUBBLES IN THE IMPLANT.
ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: D.9, H.3, H.6.
ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: H3,H6 DEVICE EVALUATION: BASED ON THE DEVICE ANALYSIS GRID, THE ASSESSMENTS OF THE COMPLAINT ARE: RUPTURE: OBSERVED, TWO OPENINGS IN POSTERIOR AND ANTERIOR SIDE ASSESSED AS SURGICAL DAMAGE. BUBBLES: UNABLE TO OBSERVE. ADDITIONAL OBSERVATIONS: CREASES ARE OBSERVED. WEAR ABRASION IS OBSERVED.
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE. HEALTHCARE PROFESSIONAL ADDITIONALLY REPORTED BUBBLES IN THE IMPLANT VIA AN RGA FORM. DEVICE HAS BEEN EXPLANTED.
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE. HEALTHCARE PROFESSIONAL ADDITIONALLY REPORTED BUBBLES IN THE IMPLANT VIA AN RGA FORM. DEVICE HAS BEEN EXPLANTED.
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE. HEALTHCARE PROFESSIONAL ADDITIONALLY REPORTED BUBBLES IN THE IMPLANT VIA AN RGA FORM. DEVICE HAS BEEN EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2704588 | INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | ALLERGAN (COSTA RICA) | SRM-405 | 3139575 | 10888628007130 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 32 YR | Female | Required Intervention |