PRIMARY PLUM SET, ORANGE POLYETHYLENE LINED LIGHT RESISTANT TUBING, CLAVE Y-SITE
Report
- Report Number
- 9615050-2022-00276
- Event Type
- Malfunction
- Date Received
- November 11, 2022
- Date of Event
- October 13, 2022
- Report Date
- October 14, 2022
- Manufacturer
- ICU MEDICAL COSTA RICA LTD.
- Product Code
- FPA
- PMA / PMN Number
- K141789
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE DEVICE WAS RECEIVED FOR EVALUATION. THE INVESTIGATION IS STILL PENDING. INITIAL REPORTER PHONE HAS AN EXTENSION NUMBER OF (B)(6).
RECEIVED ONE USED, LIST #123390488, PRIMARY PLUM SET, ORANGE POLYETHYLENE LINED LIGHT RESISTANT TUBING, CLAVE Y-SITE, SECURE LOCK, 103 INCH WITH LOT #5934274 AN ONE USED, LIST #UNKNOWN, BAG SPIKE ADAPTER WITH SPIROS WITH LOT #UNKNOWN. AS RECEIVED THE AIR FILTER ON THE VENT PLUG JUNCTURE WAS WETTED OUT WITH PRIOR INFUSATE. THE PRODUCT WAS TESTED PER PRODUCT SPECIFICATION. THERE WAS A LEAK OBSERVED FROM THE FILTER ON THE VENT PLUG JUNCTURE WITH THE CAP OPEN. THE LEAK APPEARED TO SEEP THROUGH THE FILTER VENT MATERIAL FROM VARIOUS LOCATIONS. THE REPORTED COMPLAINT CAN BE CONFIRMED. THE PROBABLE CAUSE OF LEAKAGE IS DUE TO A TEMPORARY OR COMPLETE LOSS OF HYDROPHOBIC PROPERTIES OF THE FILTER VENT MATERIAL DUE TO AN INFUSATE INTERACTION DURING USE. THE LOT HISTORY WAS REVIEWED AND NO NONCONFORMITIES WERE IDENTIFIED THAT MAY HAVE CONTRIBUTED TO THE REPORTED COMPLAINT.
THE EVENT INVOLVED A PRIMARY PLUM SET THAT WAS REPORTED TO HAVE A LEAKAGE ON FILTER VENT WITH VENT CAP CLOSED DURING INFUSION OF IFOSFAMIDE AND MESNA MEDICATION. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT HARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2878092 | PRIMARY PLUM SET, ORANGE POLYETHYLENE LINED LIGHT RESISTANT TUBING, CLAVE Y-SITE | SET, ADMINISTRATION, INTRAVASCULAR | FPA | ICU MEDICAL COSTA RICA LTD. | 1233912 | 5934274 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | IFOSFAMIDE, MFR UNK| MESNA, MFR UNK |