FDA Adverse Event Malfunction Summary report: N

RESTORE ADVANCED RECHARGEABLE

MDR report key: 1576848 · Received January 8, 2010

Report

Report Number
3004209178-2010-00180
Event Type
Malfunction
Date Received
January 8, 2010
Date of Event
December 15, 2009
Report Date
December 15, 2009
Manufacturer
MDT PUERTO RICO OPERATIONS CO., JUNCOS
Product Code
LGW
PMA / PMN Number
P840001
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B) (4)

Description of Event or Problem · 1

IT WAS REPORTED THE INTRA-OPERATIVE IMPEDANCE READINGS WERE >4000 OHMS ON ELECTRODE 4-7. ELECTRODES 0-3 WERE WITHIN NORMAL RANGE. REP REPORTED RUNNING ELECTRODE IMPEDANCE TEST WERE RUN AT THE DEFAULT VALUE. THE AMPLITUDE WAS INCREASED TO 3.0V AND RE-TESTED. ALL IMPEDANCES VALUES WERE THEN WITHIN NORMAL RANGES FROM 900-1900 OHMS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RESTORE ADVANCED RECHARGEABLE LGW MDT PUERTO RICO OPERATIONS CO., JUNCOS 37713 NA

Patients

Seq Age Sex Outcome Treatment
1 44 YR LEAD: MODEL 3890, LOT# J0331973V| EXPLANTED:| EXPLANTED:| IMPLANTED:| ADAPTER: MODEL 74002, LOT# UNKNOWN| EXTENSION: MODEL 7489, LOT# NHU012250V| IMPLANTED:| EXTENSION: MODEL 7489, LOT# NHU012247V| IMPLANTED:| IMPLANTED:| LEAD: MODEL 3890, LOT# J0331973V| EXPLANTED:| EXPLANTED:| EXPLANTED:| IMPLANTED: