FDA Adverse Event Injury Summary report: N

IMPL TWIST MP-1 3.75 MM 8 MM

MDR report key: 15764517 · Received November 9, 2022

Report

Report Number
0002023141-2022-02823
Event Type
Injury
Date Received
November 9, 2022
Date of Event
October 20, 2022
Report Date
November 9, 2022
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024013117
PMA / PMN Number
K013494
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT (B)(4). ADDITIONAL 510(K) NUMBER IS K943604. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE IMPLANT WAS REMOVED DUE TO PERI IMPLANTITIS. TOOTH SITE #14. SITE WAS GRAFTED WITH ALLOGRAFT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2951084 IMPL TWIST MP-1 3.75 MM 8 MM DENTAL IMPLANT DZE ZIMMER DENTAL 1988 2018100160 00889024013117

Patients

Seq Age Sex Outcome Treatment
1 74 YR Male Required Intervention