FDA Adverse Event Malfunction Summary report: N

XHIBIT TELEMETRY RECEIVER

MDR report key: 15760425 · Received November 9, 2022

Report

Report Number
3010157426-2022-00052
Event Type
Malfunction
Date Received
November 9, 2022
Date of Event
October 8, 2022
Report Date
October 10, 2022
Manufacturer
SPACELABS HEALTHCARE
Product Code
MHX
PMA / PMN Number
K141156
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

A SPACELABS PRODUCT SUPPORT SPECIALIST (PSS) REVIEWED XTR LOGS AND FOUND THE CHANNEL TUNED TO XTR19 AND UNABLE TO COMMUNICATE WITH A QUAD RECEIVER CARD (QRC). THE PSS DISABLED THE QRC AND MOVED THE CHANNEL TO PREVENT RECURRENCE. THE PSS WORKED WITH A HOSPITAL BIOMEDICAL ENGINEER TO EVALUATE THE TELEMETRY SYSTEM AND THE REPORTED ISSUE COULD NOT BE REPLICATED. THE QRC WAS FOUND TO BE WORKING AS EXPECTED. AT THIS TIME, THE CAUSE OF THE REPORTED ISSUE COULD NOT BE DETERMINED.

Description of Event or Problem · 0

SPACELABS HEALTHCARE WAS NOTIFIED THAT XHIBIT TELEMETRY RECEIVER (XTR) BED 6648 DISAPPEARED FROM BOTH XHIBIT CENTRAL STATIONS. THERE IS NO PATIENT OR USER HARM REPORTED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2432175 XHIBIT TELEMETRY RECEIVER TELEMETRY RECEIVER MHX SPACELABS HEALTHCARE 96280

Patients

Seq Age Sex Outcome Treatment
1 Unknown