FDA Adverse Event
Malfunction
Summary report: N
XHIBIT TELEMETRY RECEIVER
MDR report key: 15760425
·
Received November 9, 2022
Report
- Report Number
- 3010157426-2022-00052
- Event Type
- Malfunction
- Date Received
- November 9, 2022
- Date of Event
- October 8, 2022
- Report Date
- October 10, 2022
- Manufacturer
- SPACELABS HEALTHCARE
- Product Code
- MHX
- PMA / PMN Number
- K141156
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- 003
Narratives
Additional Manufacturer Narrative · 0
A SPACELABS PRODUCT SUPPORT SPECIALIST (PSS) REVIEWED XTR LOGS AND FOUND THE CHANNEL TUNED TO XTR19 AND UNABLE TO COMMUNICATE WITH A QUAD RECEIVER CARD (QRC). THE PSS DISABLED THE QRC AND MOVED THE CHANNEL TO PREVENT RECURRENCE. THE PSS WORKED WITH A HOSPITAL BIOMEDICAL ENGINEER TO EVALUATE THE TELEMETRY SYSTEM AND THE REPORTED ISSUE COULD NOT BE REPLICATED. THE QRC WAS FOUND TO BE WORKING AS EXPECTED. AT THIS TIME, THE CAUSE OF THE REPORTED ISSUE COULD NOT BE DETERMINED.
Description of Event or Problem · 0
SPACELABS HEALTHCARE WAS NOTIFIED THAT XHIBIT TELEMETRY RECEIVER (XTR) BED 6648 DISAPPEARED FROM BOTH XHIBIT CENTRAL STATIONS. THERE IS NO PATIENT OR USER HARM REPORTED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2432175 | XHIBIT TELEMETRY RECEIVER | TELEMETRY RECEIVER | MHX | SPACELABS HEALTHCARE | 96280 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |