FDA Adverse Event Malfunction Summary report: N

CLICKFINE 32GX4MM DONGBAO 7CT

MDR report key: 15760420 · Received November 9, 2022

Report

Report Number
3002806818-2022-00024
Event Type
Malfunction
Date Received
November 9, 2022
Date of Event
October 17, 2022
Report Date
November 9, 2022
Manufacturer
YPSOMED AG
Product Code
FMI
PMA / PMN Number
K171984
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE NEEDLE BROKE DURING INJECTION. A REVIEW OF PRODUCTION RECORDS SHOWED NO ABNORMALITIES OR DEVIATIONS FROM THE VALIDATED MANUFACTURING PROCESS FOR THE MAIN BATCH 201098. A RESISTANCE TO BREAKAGE TEST WAS COMPLETED ON 5 SAMPLES FROM THE SAME BATCH.ACCORDING TO ISO 9626. ALL SAMPLES PASSED THE TEST, NO BREAKAGE CAN BE FOUND.

Description of Event or Problem · 0

THE NEEDLE BROKE OFF DURING INJECTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2432170 CLICKFINE 32GX4MM DONGBAO 7CT PEN NEEDLE FMI YPSOMED AG CLICKFINE 32GX4MM DONGBAO 7CT 201098-04

Patients

Seq Age Sex Outcome Treatment
1 Unknown