FDA Adverse Event
Malfunction
Summary report: N
CLICKFINE 32GX4MM DONGBAO 7CT
MDR report key: 15760420
·
Received November 9, 2022
Report
- Report Number
- 3002806818-2022-00024
- Event Type
- Malfunction
- Date Received
- November 9, 2022
- Date of Event
- October 17, 2022
- Report Date
- November 9, 2022
- Manufacturer
- YPSOMED AG
- Product Code
- FMI
- PMA / PMN Number
- K171984
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- 003
Narratives
Additional Manufacturer Narrative · 0
THE NEEDLE BROKE DURING INJECTION. A REVIEW OF PRODUCTION RECORDS SHOWED NO ABNORMALITIES OR DEVIATIONS FROM THE VALIDATED MANUFACTURING PROCESS FOR THE MAIN BATCH 201098. A RESISTANCE TO BREAKAGE TEST WAS COMPLETED ON 5 SAMPLES FROM THE SAME BATCH.ACCORDING TO ISO 9626. ALL SAMPLES PASSED THE TEST, NO BREAKAGE CAN BE FOUND.
Description of Event or Problem · 0
THE NEEDLE BROKE OFF DURING INJECTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2432170 | CLICKFINE 32GX4MM DONGBAO 7CT | PEN NEEDLE | FMI | YPSOMED AG | CLICKFINE 32GX4MM DONGBAO 7CT | 201098-04 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |