FDA Adverse Event Malfunction Summary report: N

ADVIA CENTAUR SARS-COV-2 IGG (SCOVG)

MDR report key: 15756867 · Received November 8, 2022

Report

Report Number
1219913-2022-00389
Event Type
Malfunction
Date Received
November 8, 2022
Date of Event
October 20, 2022
Report Date
December 22, 2022
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS, INC.
Product Code
QKO
PMA / PMN Number
EUA202670
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HU
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MDR 1219913-2022-00389 WAS INITIALLY FILED ON 2022-11-08. ADDITIONAL INFORMATION, 2022-12-13: SIEMENS HEALTHCARE DIAGNOSTICS INC. INVESTIGATION CONFIRMED THAT THE FAILED CALIBRATIONS ARE DUE TO THE STANDARD S01 VALUE IN THE MASTER CURVE CARD (MCC) FOR THE ADVIA CENTAUR SYSTEMS LOTS 19529015 AND 27236015, BEING SET AT 0.1 INDEX. THE S01 VALUE SHOULD BE 0.01 INDEX. THE ISSUE IS ISOLATED TO THE ADVIA CENTAUR SYSTEMS SCOVG LOTS 19529015 AND 27236015, SMNS (B)(4). CUSTOMERS WERE NOTIFIED OF THE ISSUE AND INSTRUCTED TO DISCONTINUE USE OF THE KITS. (US: CC 23-01.A.US WAS SENT TO US CUSTOMERS ON (B)(6) 2022 AND OUTSIDE THE US: CC 23-01.A.OUS WAS SENT TO CUSTOMERS ON (B)(6) 2022 WHO HAVE RECEIVED THE ADVIA CENTAUR SYSTEMS SARS-COV-2 IGG (SCOVG) LOTS 19529015 AND 27236015). NOTE: IN SECTION H6, THE CODES FOR TYPE OF INVESTIGATION, INVESTIGATION FINDINGS AND INVESTIGATION CONCLUSION HAVE BEEN UPDATED.

Additional Manufacturer Narrative · 0

A CUSTOMER FROM OUTSIDE OF THE UNITED STATES OBSERVED REPRODUCIBLE CALIBRATION FAILURE DUE TO EXCESSIVE DEVIATION FOR THE LOW CALIBRATOR WHEN USING THE ADVIA CENTAUR XPT SARS-COV-2 IGG (SCOVG) ASSAY. THERE ARE NO ALLEGATIONS OF PATIENT HARM, CHANGES IN TREATMENT, OR DIAGNOSTIC DELAYS DUE TO THE OBSERVED CALIBRATION FAILURE. SIEMENS HEALTHCARE DIAGNOSTICS IS INVESTIGATING THE OBSERVED PROBLEM. ALTHOUGH THERE IS NO POTENTIAL FOR SERIOUS INJURY IN THIS CASE, AN MDR WILL BE REPORTED TO THE FDA AS A REQUIREMENT OF THE EMERGENCY USE AUTHORIZATION (EUA).

Description of Event or Problem · 0

THE CUSTOMER REPORTS OBSERVATION OF REPRODUCIBLE CALIBRATION FAILURE DUE TO EXCESSIVE DEVIATION FOR THE LOW CALIBRATOR WHEN USING THE ADVIA CENTAUR SARS-COV-2 IGG (SCOVG) ASSAY. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE SCOVG CALIBRATION FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2954071 ADVIA CENTAUR SARS-COV-2 IGG (SCOVG) SARS-COV-2 IMMUNOASSAY QKO SIEMENS HEALTHCARE DIAGNOSTICS, INC. N/A 015