FDA Adverse Event Injury Summary report: N

HOYA 255

MDR report key: 15756455 · Received November 8, 2022

Report

Report Number
3006723646-2022-00157
Event Type
Injury
Date Received
November 8, 2022
Date of Event
October 26, 2022
Report Date
October 27, 2022
Manufacturer
HOYA SURGICAL OPTICS, INC.
Product Code
HQL
PMA / PMN Number
P080004
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS INITIAL REPORT IS BEING SUBMITTED TO FDA FOR A REPORTABLE EVENT THAT OCCURRED OUTSIDE THE USA. HAPTIC DAMAGE IS INDICATED AS A POTENTIAL MALFUNCTION RELATED TO THE IOL, AS COVERED UNDER THE WARNINGS SECTION OF THE PRODUCT'S INSTRUCTIONS FOR USE (IFU). MANUFACTURER'S CODES FOR: TYPE OF INVESTIGATION, FINDINGS, AND CONCLUSION ARE PENDING COMPLETION OF PRODUCT RETURN AND INVESTIGATION. ONCE THE PRODUCT INVESTIGATION IS COMPLETED, A FOLLOW-UP REPORT WILL BE SUBMITTED TO FDA WHICH WILL INCLUDE THE MANUFACTURER'S CODES FOR TYPE OF INVESTIGATION, FINDINGS, AND CONCLUSION.

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP #1 EMDR IS BEING SUBMITTED TO FDA FOR A REPORTABLE EVENT THAT OCCURRED OUTSIDE OF THE USA. THE REPORT INCLUDES CORRECTED INFORMATION AND ADDITIONAL INFORMATION NOT AVAILABLE/INCLUDED IN THE INITIAL REPORT. PORTIONS OF THE PRODUCT WERE RETURNED TO THE MANUFACTURER. THE INVESTIGATION WAS CONDUCTED, WITH THE METHODS AND RESULTS AS NOTED BELOW. THE IOL WAS AVAILABLE FOR INVESTIGATION. APPEARANCE CHECK RESULT WAS CONSISTENT TO REPORTED INFORMATION. NO ABNORMALITIES WERE FOUND IN PRODUCTION AND INSPECTION RECORDS OF THE PRODUCT. (SERIAL NO.: (B)(6) ; MODEL: 255). THE EXACT ROOT CAUSE OF THE EVENT WAS NOT DETERMINED. HOWEVER, BASED ON AVAILABLE INFORMATION, WE BELIEVE THIS EVENT WAS NOT CAUSED BY OUR PRODUCT QUALITY. CAPA-22-0009 HAS BEEN INITIATED FOR "DAMAGED HAPTIC" COMPLAINTS.

Description of Event or Problem · 0

DAMAGED HAPTIC AFTER IMPLANTATION. THE DOCTOR EXPLANTED THE IOL FROM THE EYE SINCE HE NOTICED THE TRAILING HAPTIC WAS SEPARATED. THE SURGERY WAS COMPLETED WITH ANOTHER IOL.

Description of Event or Problem · 0

DAMAGED HAPTIC AFTER IMPLANTATION. THE DOCTOR EXPLANTED THE IOL FROM THE EYE SINCE HE NOTICED THE TRAILING HAPTIC WAS SEPARATED. THE SURGERY WAS COMPLETED WITH ANOTHER IOL

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2875467 HOYA 255 INTRAOCULAR LENS HQL HOYA SURGICAL OPTICS, INC. HOYA 255 (+24.50 D)

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other